The comparison of absorbed dose measurements for water and artificial body fluid

نویسندگان

  • A. Arasoglu
  • A. Gur
چکیده مقاله:

Background: Advances in radiation dosimetry concepts and the development of primary measurement standards based on absorbed dose to water over the last decades offer the possibility to calibrate ionisation chambers directly in terms of absorbed dose to water. The aim of this study is the investigation on utility of artificial body fluid (ABF) instead of water by comparing dosimetric measurements for radiotherapy between water and ABF which is more close to human tissue. Materials and Methods: The measurements were done using 60Co gamma source with a radiation field sizes of 5×5, 10×10, 15×15, 20×20 and 25×25 cm2 at PTW Freiburg MP3 water phantom front surface. The comparisons of the dose measurements were obtained by using IAEA TRS-398 dosimetry protocols and Mephysto mc2 dose analyzer program. Percent depth dose (PDD), dose profiles and penumbras are compared for water and ABF. Results: When the results of the PDD for water and ABF were compared, the maximum difference was observed in big field sizes. The difference in penumbras was found 2.3 mm averagely for depth of maximum dose (dmax). In addition same differences were observed between water and ABF when the dose profiles were compared. It is found that PDD values taken for water are good agreed with PDD values published in British Journal of Radiology (BJR) Supplement 25. Conclusion: Since the ABF is more equivalent to human tissue than water, it is suggested that advanced dosimetric studies should be performed with ABF instead of water. Iran. J. Radiat. Res., 2012 10(3‐4): 157‐164

برای دانلود باید عضویت طلایی داشته باشید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

the comparison of absorbed dose measurements for water and artificial body fluid

background: advances in radiation dosimetry concepts and the development of primary measurement standards based on absorbed dose to water over the last decades offer the possibility to calibrate ionisation chambers directly in terms of absorbed dose to water. the aim of this study is the investigation on utility of artificial body fluid (abf) instead of water by comparing dosimetric measurement...

متن کامل

the evaluation and comparison of two esp textbooks available on the iranian market for teaching english to the students of medicine

abstract this study evaluated and compared medical terminology and english for the students of medicine (ii) as two representatives of the textbooks available on the iranian market for teaching english to the students of medicine. this research was performed on the basis of a teacher’s and a number of students’ attitudes and the students’ needs analysis for two reasons: first, to investigate...

15 صفحه اول

Absorbed dose to water reference dosimetry using solid phantoms in the context of absorbed-dose protocols.

For reasons of phantom material reproducibility, the absorbed dose protocols of the American Association of Physicists in Medicine (AAPM) (TG-51) and the International Atomic Energy Agency (IAEA) (TRS-398) have made the use of liquid water as a phantom material for reference dosimetry mandatory. In this work we provide a formal framework for the measurement of absorbed dose to water using ioniz...

متن کامل

Uncertainties on measurements of absorbed-dose-to-water, from clinical high- energy eletron beams: a comparison with the IAEA protocols

A pilot study was performed to verify the troubles involved in the implementation of dosimetric protocol TRS 398 for absorbed-dose-to-water, using clinical high-energy electron beams and to accomplish an detailed evaluation of uncertainty chain components associated to the measurement, which ones were not treated in the IAEA protocols, aimed the standardization of technical procedures.

متن کامل

on the comparison of keyword and semantic-context methods of learning new vocabulary meaning

the rationale behind the present study is that particular learning strategies produce more effective results when applied together. the present study tried to investigate the efficiency of the semantic-context strategy alone with a technique called, keyword method. to clarify the point, the current study seeked to find answer to the following question: are the keyword and semantic-context metho...

15 صفحه اول

منابع من

با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

ذخیره در منابع من قبلا به منابع من ذحیره شده

{@ msg_add @}


عنوان ژورنال

دوره 10  شماره None

صفحات  157- 164

تاریخ انتشار 2012-12

با دنبال کردن یک ژورنال هنگامی که شماره جدید این ژورنال منتشر می شود به شما از طریق ایمیل اطلاع داده می شود.

کلمات کلیدی

میزبانی شده توسط پلتفرم ابری doprax.com

copyright © 2015-2023