Physical and Dosimetric Aspect of Euromechanics Add-on Multileaf Collimator on Varian Clinac 2100 C/D

نویسندگان

  • A Mostaar Department of Medical Physics, School of Medicine, Shahid Beheshti of Medical Science, Tehran, Iran
  • Gh Geraily Department of Medical Physics, Tehran University of Medical Science, Tehran, Iran
  • R Mohammadi Department of Medical Physics, Iran University of Medical Science, Tehran, Iran
  • S A Rohani Department of Medical Physics, Tehran University of Medical Science, Tehran, Iran
  • S R Mahdavi Radiation biology research center & medical Physics department, faculty of medicine, Iran University of Medical Science, Tehran, Iran
  • S Ueltzhöffer Department of Clinic for Radiotherapy and RadioOncology, Medical Faculty Mannheim of the University of Heidelberg, Heidelberg, Germany
چکیده مقاله:

Background: Before treatment planning and dose delivery, quality assurance of multi-leaf collimator (MLC) has an important role in intensity-modulated radiation therapy (IMRT) due to the creation of multiple segments from optimization process.Objective: The purpose of this study is to assess the quality control of MLC leaves using EBT3 Gafchromic films.Material and Methods: Leaf Position accuracy and leaf gap reproducibility were checked with Garden fence test. The garden fence test consists of 5 thin bands A) 0.2 Cm width spaced at 2 Cm intervals and B) 1 Cm width spaced at 1 Cm intervals. Each leaf accuracy was analyzed with measuring the full-width half-maximum (FWHM). Maximum and average leaf transmission were measured with gafchromic EBT3 films from Ashland for both 6 MV and 18 MV beams.Results: Leaf positions were found to be in a range between 1.78 – 2.53 mm, instead of nominal 2 mm for the test A and between 9.09 – 10.36 mm, instead of nominal 10 mm for the test B. The Average radiation transmission of the MLC was noted 1.79% and 1.98% of the open 10x10 Cm2 field at isocenter for 6 MV and 18 MV beams, respectively. Maximum radiation transmission was noted 4.1% and 4.4% for 6 MV and 18 MV beams, respectively.Conclusion: In this study, application of gafchromic EBT3 films for the quality assurance of Euromechanics multileaf collimator was studied. Our results showed that the average leaf leakage and positional accuracy of this type of MLC were in the acceptance level based on the Protocols.

برای دانلود باید عضویت طلایی داشته باشید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Beam characteristic of P.E collimators Add-on multileaf collimator

  Introduction:    Multileaf collimators (MLCs) are computer-controlled devices which are Useful and cost-effective tools for conformal therapy and intensity modulation radiotherapy (IMRT). Nowadays MLCs are important and standard configuration of the new medical linear accelerators which are substitute of conventional Blocks and has eliminated the ...

متن کامل

The Dosimetric Effects of Different Multileaf Collimator Widths on Physical Dose Distributions

Introduction: Geometric changes in the multileaf collimator (MLC) led to dosimetric considerations in intensity-modulated radiation therapy (IMRT) due to the number and size of the pixels in the intensity map, which are determined by the MLC leaf width. In this study, we evaluated the dosimetric effects of different MLC widths on physical dose distributions for IMRT plans. Materials and Method...

متن کامل

The Dosimetric Effects of Different Multileaf Collimator Widths on Physical Dose Distributions

Article type: Original Article Introduction: Geometric changes in the multileaf collimator (MLC) led to dosimetric considerations in intensity-modulated radiation therapy (IMRT) due to the number and size of the pixels in the intensity map, which are determined by the MLC leaf width. In this study, we evaluated the dosimetric effects of different MLC widths on physical dose distributions for IM...

متن کامل

Physical and dosimetric characteristic of high‐definition multileaf collimator (HDMLC) for SRS and IMRT

Physical and dosimetric characteristics of HDMLC were studied for SRS6, 6, and 10 MV X-rays from Novalis Tx. This in-built tertiary collimator consists of 60 pairs (32 × 0.25 cm; 26 × 0.5 cm and 2 × 0.7 cm) of leaves. Properties of HDMLC studied included alignment, readout and radiation field congruence, radiation penumbra, accuracy and reproducibility of leaf position and gap width, static and...

متن کامل

Monte Carlo modeling of a Novalis TX Varian 6 MV with HD‐120 multileaf collimator

A Monte Carlo model of the Novalis Tx linear accelerator equipped with high-definition multileaf collimator (HD-120 HD-MLC) was commissioned using ionization chamber measurements in water. All measurements in water were performed using a liquid filled ionization chamber. Film measurements were made using EDR2 film in solid water. Open rectangular fields defined by the jaws or the HD-MLC were us...

متن کامل

Impact of Multileaf Collimator Configuration Parameters on the Dosimetric Accuracy of 6-MV Intensity-Modulated Radiation Therapy Treatment Plans

The purpose of this study was to evaluate the impact of selected configuration parameters that govern multileaf collimator (MLC) transmission and rounded leaf offset in a commercial treatment planning system (TPS) (Pinnacle3, Philips Medical Systems, Andover, MA, USA) on the accuracy of intensity-modulated radiation therapy (IMRT) dose calculation. The MLC leaf transmission factor was modified ...

متن کامل

منابع من

با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

ذخیره در منابع من قبلا به منابع من ذحیره شده

{@ msg_add @}


عنوان ژورنال

دوره 9  شماره 1

صفحات  -

تاریخ انتشار 2019-02-01

با دنبال کردن یک ژورنال هنگامی که شماره جدید این ژورنال منتشر می شود به شما از طریق ایمیل اطلاع داده می شود.

میزبانی شده توسط پلتفرم ابری doprax.com

copyright © 2015-2023