Organ Dose Measurement in Computed Tomography Using Thermoluminescence Dosimeter in Locally Developed Phantoms

نویسندگان

  • Aderonke Adedokun Department of Radiation Biology and Radiotherapy, College of Medicine, University of Lagos.
  • AKINTAYO OMOJOLA Department of Radiology, Medical Physics Unit, Federal medical Centre, Asaba.
  • Dare Adewa Department of radiation biology, radiotherapy, radiodiagnosis, Faculty of clinical sciences, university of lagos, nigeria
  • Moses Aweda Department of Radiation Biology, Radiotherapy, Radiodiagnosis and Radiography, College of Medicine, Lagos University Teaching Hospital, Idi-Araba, Lagos, Nigeria
  • Muhammad Habeebu Department of Radiation Biology, Radiotherapy, Radiodiagnosis and Radiography, College of Medicine, Lagos University Teaching Hospital, Idi-Araba, Lagos, Nigeria
  • Nusirat Adedewe Department of Radiotion biology, radiotherapy, radiodiagnosis, Faculty of clinical sciences, University of Lagos.
  • SAMUEL ADENEYE Department of Radiation Biology, Radiotherapy, Radiodiagnosis and Radiography, College of Medicine, Lagos University Teaching Hospital, Idi-Araba, Lagos, Nigeria
  • SUNDAY AJIBADE Radiation Biology, Radiotherapy, Radiodiagnosis, and Radiography, Clinical Sciences, College of Medicine, University of Lagos, Lagos State, Nigeria
  • Victor Ekpo Department of radiation biology, radiotherapy, radiodiagnosis, faculty of clinical sciences, university of lagos, nigeria
چکیده مقاله:

Introduction: Organ dose estimation using thermoluminescence dosimeter (TLD) is known to be a standard, although many other methods, such as simulation software, optically stimulated luminescent dosimeters, and photodiodes are still in use. This study aimed at directly measuring mean organ doses to the selected organs in the head/neck, chest, and abdominal regions from four computed tomography (CT) units in Lagos, south-west of Nigeria. Material and Methods: This study was conducted on locally constructed inhomogeneous phantoms to measure mean organ doses to the head/neck, chest, and abdominopelvic regions from CT units in the Lagos metropolis, Nigeria. Lithium fluoride doped with magnesium and titanium (LiF: Mg, Ti) TLD was used for the measurement. Statistical analysis was performed by IBM SPSS (version 20). Results: Validation of the designed phantoms was below ± 20% kVp and mAs parameters among the CT units, which was statistically different with regard to the observed dose discrepancies. Generally, a one-way ANOVA showed that there was a statistically significant difference in the investigated mean organ dose (P = 0.043). The comparison of the obtained results from this study with those of other studies revealed that there was no statistically significant difference in the TLDs (P > 0.05). The maximum relative difference in the dose was < 200%. Conclusion: The designed phantoms seemed to be useful for CT dose validation and could be used to validate simulation software in areas where readymade phantoms are not available.

برای دانلود باید عضویت طلایی داشته باشید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Thorax organ dose estimation in computed tomography based on patient CT data using Monte Carlo simulation

Background: &nbsp;This study presents patient specific and organ dose estimation in computed tomography (CT) imaging of thorax directly from patient CT image using Monte Carlo simulation.&nbsp; Patient's CT image is considered as the patient specific phantom and the best representative of patient physical index in order to calculate specific organ dose. Materials and Methods: EGSnrc /BEAMnr...

متن کامل

Dose Reduction Methods in Chest Computed Tomography

Introduction: Computed tomography is one of the most powerful tools for investigating thorax disease, because it shows progression of lung disease much clearer than chest radiography. Technical advances in CT scanners have led to some abilities for CT scanners such as coverage of larger volume, lower noise and ability for acquiring image in one respiratory phase. However, this will be accompani...

متن کامل

Evaluation of Maximum Patient Skin Dose Arising from Interventional Cardiology Using Thermoluminescence Dosimeter in Mashhad, Iran

Introduction The increasing practice of interventional fluoroscopy in diagnosis and treatment of cardiovascular disease has risen attention to improve radiation protection of patients and cardiologists in these relatively high dose techniques. Therefore, nowadays there is an emphasis on the measurement of radiation dose received by patients and cardiologists arising from the relevant procedures...

متن کامل

thorax organ dose estimation in computed tomography based on patient ct data using monte carlo simulation

background:  this study presents patient specific and organ dose estimation in computed tomography (ct) imaging of thorax directly from patient ct image using monte carlo simulation.  patient&apos;s ct image is considered as the patient specific phantom and the best representative of patient physical index in order to calculate specific organ dose. materials and methods: egsnrc /beamnrc monte c...

متن کامل

Organ point dose measurements in clinical multi slice computed tomography (MSCT) examinations with the MOSkinTM radiation dosimeter

This study reports on the application of the MOSkinTM dosimeter in MSCT imaging for the real-time measurement of absorbed organ point doses in a tissue-equivalent female anthropomorphic phantom. MOSkinTM dosimeters were placed within the phantom to measure absorbed point organ doses for 2 commonly applied clinical scan protocols, namely the renal calculus scan and the pulmonary embolus scan. Me...

متن کامل

Implications of Patient Centring on Organ Dose in Computed Tomography.

Automatic exposure control (AEC) in computed tomography (CT) facilitates optimisation of dose absorbed by the patient. The use of AEC requires appropriate 'patient centring' within the gantry, since positioning the patient off-centre may affect both image quality and absorbed dose. The aim of this experimental study was to measure the variation in organ and abdominal surface dose during CT exam...

متن کامل

منابع من

با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

ذخیره در منابع من قبلا به منابع من ذحیره شده

{@ msg_add @}


عنوان ژورنال

دوره 16  شماره 2

صفحات  126- 132

تاریخ انتشار 2019-03-01

با دنبال کردن یک ژورنال هنگامی که شماره جدید این ژورنال منتشر می شود به شما از طریق ایمیل اطلاع داده می شود.

میزبانی شده توسط پلتفرم ابری doprax.com

copyright © 2015-2023