Effects of Kilovoltage on Image Quality and Entrance Surface Dose in Lumbar Spine Digital Radiography
نویسندگان
چکیده مقاله:
Introduction: Digital radiography possesses a wide dynamic range and has a major advantage in producing an acceptable image of diagnostic value even though overexposure occurs. Lumbar spine (LS) radiography is the most common examinations that gives high radiation dose to patients and accounts for the highest collective population dose of any conventional radiographic examinations. As such, this study was carried out to ascertain the impact of image quality and entrance surface dose (ESD) with different exposure settings in the anteroposterior (AP) and lateral LS. Material and Methods: The torso of the PBU-50 phantom was exposed to medium and high kilovoltage peak (kVp). A total of 14 images for LS were obtained. Relative image quality was assessed using Leeds Test Objects TOR CDR whilst the ESD was ascertained using an optically stimulated luminescence dosimeter. Results: The results of Friedman test indicated a significant difference in image quality when using medium and high kVp. Wilcoxon signed-rank test also reflected a significant difference in ESD between the use of medium and high kVp for both AP and lateral LS. Conclusion: Significant differences in image quality and ESD were obtained using medium and high kVp with medium kVp resulting in high contrast but low contrast sensitivity and vice versa. The findings of the present study indicated that the recommended kVp for AP LS was from 75kVp to 81kVp whilst for lateral LS the recommended kVp was from 85kVp to 90kVp for an average adult patient.
منابع مشابه
Investigation of the Entrance Surface Dose and Dose to Different Organs in Lumbar Spine Imaging
Background: Dose assessment using proper dosimeters is especially important in radiation protection optimization and imaging justification in diagnostic radiology.Objective: The aim of this study is to obtain the Entrance Skin Dose (ESD) of patients undergoing lumbar spine imaging using two thermoluminescence dosimeters TLD-100 (LiF: Mg, Ti) and GR-200 (LiF: Mg, Cu, P) and also to obtain the ab...
متن کاملImage Quality and Entrance Surface Dose Evaluation of Lateral Cervical Spine: A Study Using Grid and Non-Grid Techniques
Introduction: The purpose of this study is to investigate the effects of grid and non-grid techniques in the lateral cervical spine radiography on image quality and entrance surface dose (ESD). Although image quality and radiation doses have been studied by researchers, there is still a dearth of information on image quality and patient dose with different techniques. Material and Methods: The...
متن کاملinvestigation of the entrance surface dose and dose to different organs in lumbar spine imaging
background: dose assessment using proper dosimeters is especially important in radiation protection optimization and imaging justification in diagnostic radiology. objective: the aim of this study is to obtain the entrance skin dose (esd) of patients undergoing lumbar spine imaging using two thermoluminescence dosimeters tld-100 (lif: mg, ti) and gr-200 (lif: mg, cu, p) and also to obtain the a...
متن کاملThe Practice of Chest Radiography Using Different Digital Imaging Systems: Dose and Image Quality
Introduction: The study was undertaken to evaluate the practice of chest radiography using different digital imaging systems and its influence on dose and image quality. Materials and Methods: The study was carried out in two hospitals from March 2016 to June 2016. Sixty ambulatory patients aged 21 to 60 years who were able to cooperate without difficulty and weighed between 60 to 80 kg were ...
متن کاملInvestigation of the Entrance Surface Dose and Dose to Different Organs in Lumbar Spine Imaging
BACKGROUND Dose assessment using proper dosimeters is especially important in radiation protection optimization and imaging justification in diagnostic radiology. OBJECTIVE The aim of this study is to obtain the Entrance Skin Dose (ESD) of patients undergoing lumbar spine imaging using two thermoluminescence dosimeters TLD-100 (LiF: Mg, Ti) and GR-200 (LiF: Mg, Cu, P) and also to obtain the a...
متن کاملEffects of Focus Film Distance (ffd) Variation on Entrance Testicular Dose in Lumbar-pelvic Radiography
INTRODUCTION With the steady increase in public and professional concern regarding the biological effects of ionising radiation, there is a need for both the Chiropractic and Radiography professions to improve imaging techniques for the lowering of patient radiation doses. Lumbar radiographs are essential in chiropractic general practice for biomechanical diagnosis and postural analysis. Detail...
متن کاملمنابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
ذخیره در منابع من قبلا به منابع من ذحیره شده{@ msg_add @}
عنوان ژورنال
دوره 16 شماره 5
صفحات 368- 371
تاریخ انتشار 2019-09-01
با دنبال کردن یک ژورنال هنگامی که شماره جدید این ژورنال منتشر می شود به شما از طریق ایمیل اطلاع داده می شود.
کلمات کلیدی
میزبانی شده توسط پلتفرم ابری doprax.com
copyright © 2015-2023