Dosimetric characteristics of 137Cs sources used in after loading Selectron system by Monte Carlo method
نویسندگان
چکیده مقاله:
Background: For an effective treatment planning in brachytherapy, it is necessary to know the accurate source dosimetric information such as air kerma strength, exposure rate constant, dose rate constant and redial dose distribution. The usual method to determine these factors is thermo luminescent dosimeter (TLD) dosimetry. Nowadays, another more accurate method is known to be the Monte Carlo simulation, and it is widely used in determining of brachytherapy sources dosimetric factors. The aim of the present study has been determine to dose distribution of 137Cs sources configuration used in interacavitary low dose rate (LDR) Selectron brachytherapy by Monte Carlo simulation method. Materials and Methods: Simulation has been performed for 137Cs stainless-steel encapsulated source in 2.5 mm in diameter of Selectron system. Six different combinations of 137Cs sources were simulated and each combination was put into different applicator. The Monte Carlo MCNP4C code was used for simulation. The number of history was taken 1 million for the calculation of air kerma strength and exposure rate constant and 0.1 million for calculating dose rate constant and redials dose distribution. Results: Maximum air kerma strength was obtained for the combination with fourteen active sources and found to be 919.3 |µ|Gym2h-1 while its minimum value was 133.9 µ|Gym2h-1 for the case with two active sources. Maximum and minimum dose rate constant of 4.3 cGyh-1mCi-1 and 2.1 cGyh-1mCi-1 were obtained for the combination with eight and fourteen active sources, respectively. Conclusion: The results indicated that the Monte Carlo simulation method may be useful to determine dosimetric characteristics of brachytherapy sources.
منابع مشابه
dosimetric characteristics of 137cs sources used in after loading selectron system by monte carlo method
background: for an effective treatment planning in brachytherapy, it is necessary to know the accurate source dosimetric information such as air kerma strength, exposure rate constant, dose rate constant and redial dose distribution. the usual method to determine these factors is thermo luminescent dosimeter (tld) dosimetry. nowadays, another more accurate method is known to be the monte carlo ...
متن کاملDosimetric Characteristics of the Brachytherapy Sources Based on Monte Carlo Method
The term of brachytherapy, also known as “internal radiotherapy, sealed source radiotherapy, curietherapy or endocurietherapy” is from Greek work brachy means short distance and therapy (treatment) and also known as internal radiotherapy or sealed source radiotherapy. Brachytherapy is a special form of radiotherapy where a radioactive source is carefully placed on or inside the area to be treat...
متن کاملA Monte Carlo simulation and dosimetric verification of physical wedges used in radiation therapy
Background: The presence of a wedge filter in the beam trajectory can modify the beam quality and cause some changes in the dosimetry parameters which are usually difficult to be measured directly and accurately. Material and Methods: In this study the MCNP-4C Monte Carlo code was used to simulate the 9 MV photon beam generated by a linear accelerator. Upon getting a good agreement bet...
متن کاملCalculating Reaction Rate of Positron Emitters During Proton Therapy Which Are Used In Online PET Scan by Monte Carlo Method
Introduction: Nowadays, cancer is one of the most important concerns in human being's societies, more than one-third of people in around the world will get cancer during their lives. Physical dose depth distribution characteristic of protons in a tissue is determined with a low dose in the entrance region, maximum dose in the Bragg region, and rapid decline takes place near the...
متن کاملDosimetric investigation of high dose rate Ir-192 source with Monte Carlo method
Background: This study aims to calculate the air-kerma strength (SK), the dose rate constant (∧) and the dose rate profiles of a Gammamed 12i Ir-192 source by using the Monte Carlo technique and to compare the dose rate values with those calculated by the Abacus HDR treatment planning system (TPS). Materials and Methods: Air-kerma strength (in units of U; 1 U= μGy m2 h-1) and the dose ra...
متن کاملa monte carlo simulation and dosimetric verification of physical wedges used in radiation therapy
background: the presence of a wedge filter in the beam trajectory can modify the beam quality and cause some changes in the dosimetry parameters which are usually difficult to be measured directly and accurately. material and methods: in this study the mcnp-4c monte carlo code was used to simulate the 9 mv photon beam generated by a linear accelerator. upon getting a good agreement between the ...
متن کاملمنابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
ذخیره در منابع من قبلا به منابع من ذحیره شده{@ msg_add @}
عنوان ژورنال
دوره 5 شماره None
صفحات 147- 151
تاریخ انتشار 2007-12
با دنبال کردن یک ژورنال هنگامی که شماره جدید این ژورنال منتشر می شود به شما از طریق ایمیل اطلاع داده می شود.
میزبانی شده توسط پلتفرم ابری doprax.com
copyright © 2015-2023