Design and Evaluation of an Articulated Ankle Foot Orthosis with Plantarflexion Resistance on the Gait: a Case Series of 2 Patients with Hemiplegia

نویسندگان

  • A Daryabor PhD, Department of Orthotics and Prosthetics, University of Social Welfare and Rehabilitation Sciences, Tehran, Iran |PhD, Researcher in International University of Health and Welfare, Japan, Tokyo
  • G Aminian PhD, Department of Orthotics and Prosthetics, University of Social Welfare and Rehabilitation Sciences, Tehran, Iran
  • M Arazpour PhD, Department of Orthotics and Prosthetics, University of Social Welfare and Rehabilitation Sciences, Tehran, Iran
  • M Baniasad PhD, Mechanical Engineering Department, Sharif University of Technology, Tehran, Iran
  • S Yamamoto PhD, International University of Health and Welfare, Tokyo, Japan
چکیده مقاله:

Ankle-foot orthoses (AFOs) have been described to have positive effects on the gait biomechanics in stroke patients. The plantarflexion resistance of an AFO is considered important for hemiplegic patients, but the evidence is still limited. The purpose of this case series was to design and evaluate the immediate effect of an articulated AFO on kinematics and kinetics of lower-limb joints in stroke patients. The articulated AFO with the adjustment of plantarflexion resistance was designed. The spring generates a plantarflexion resistance of the ankle joint at initial stance phase. The efficacy of orthosis was evaluated on two stroke patients in 2 conditions: without an AFO and with the AFO. Results showed the immediate improvements for walking speed, stride length and angular changes of dorsiflexion of the paretic ankle joint during a gait cycle of both subjects using the AFO compared with barefoot walking. The AFO also was able to reduce the paretic knee extension in the single-support phase of the stance and increase the vertical COM displacement during stance phase on the affected leg. In conclusion, the designed AFO affect not only the movement of the ankle joint but also the movements of the knee joint and the vertical COM height. These changes indicate improvement of the first and the second rockers and swing phase gait but not third rocker function. Further investigation is recently underway to compare its effect compared with other AFOs on the gait parameters of hemiplegic patients. 

برای دانلود باید عضویت طلایی داشته باشید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

the evaluation of language related engagment and task related engagment with the purpose of investigating the effect of metatalk and task typology

abstract while task-based instruction is considered as the most effective way to learn a language in the related literature, it is oversimplified on various grounds. different variables may affect how students are engaged with not only the language but also with the task itself. the present study was conducted to investigate language and task related engagement on the basis of the task typolog...

15 صفحه اول

Design and impact of the ankle foot orthosis on gait for the hemiplegic patients

INTRODUCTION: Two-thirds of those suffering a cerebral vascular accident (CVA) survive, with approximately half of those being left with physical incapacity [1]. Hemiplegic patients with cerebral vascular lesion, present a clinical picture which combines a variety of motor, sensory, visual and cognitive defects. These defects typically lead to disturbances of the gait cycle, characterised by ch...

متن کامل

azerbaijans political development after the collapse of soviet union and implication on relation with the islamic republic of iran

در فصل اول این پایاین نامه در خصوص تاریخ کشور اذربایجان قبل و بعد از جدایی این کشور از ایران مورد بررسی قرار گرفته است ودر فصل دوم تحولات سیاسی این کشور بعد از 1991 و در واقع بعد از فروپاشی شوروی و دولتهایی که روی کار امدند در از جمله دولت ابولفضل ایلچی بیگ،دولت حیدر علی اف وبعد از او پسرش الهام علی اف و نگرش هرکدام از این دولتها به سیاسیت خارجی اذربایجان مورد اشاره قرار گرفته است.در ودر فصل سو...

Gait improvement of hemiplegic patients using an ankle-foot orthosis with assistance of heel rocker function.

The ankle-foot orthosis (AFO) with oil damper, GaitSolution (GS), was previously developed to make heel rocker function possible during the gait of hemiplegic patients. To clarify the characteristics of patients who show adaptation to GS use, a questionnaire was completed by the clinicians responsible for 99 patients with hemiplegia who participated in this study. Clinicians completed items con...

متن کامل

منابع من

با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

ذخیره در منابع من قبلا به منابع من ذحیره شده

{@ msg_add @}


عنوان ژورنال

دوره 10  شماره 1

صفحات  119- 128

تاریخ انتشار 2020-02-01

با دنبال کردن یک ژورنال هنگامی که شماره جدید این ژورنال منتشر می شود به شما از طریق ایمیل اطلاع داده می شود.

میزبانی شده توسط پلتفرم ابری doprax.com

copyright © 2015-2023