Comparisons of Hounsfield Unit Linearity between Images Reconstructed using an Adaptive Iterative Dose Reduction (AIDR) and a Filter Back-Projection (FBP) Techniques

نویسندگان

  • C Anam PhD, Department of Physics, Faculty of Sciences and Mathematics, Diponegoro University, Indonesia
  • H Sutanto PhD, Department of Physics, Faculty of Sciences and Mathematics, Diponegoro University, Indonesia
  • I Suyudi BSc, Department of Physics, Faculty of Sciences and Mathematics, Diponegoro University, Indonesia
  • P Triadyaksa PhD, Department of Physics, Faculty of Sciences and Mathematics, Diponegoro University, Indonesia
  • T Fujibuchi PhD, Department of Health Sciences, Faculty of Medical Sciences, Kyushu University, Japan
چکیده مقاله:

Background: The HU linearity is an essential parameter in a quantitative imaging and the treatment planning systems of radiotherapy. Objective: This study aims to evaluate the linearity of Hounsfield unit (HU) in applying the adaptive iterative dose reduction (AIDR) on CT scanner and its comparison to the filtered back-projection (FBP).Material and Methods: In this experimental phantom study, a TOS-phantom was scanned using a Toshiba Alexion 6 CT scanner. The images were reconstructed using the FBP and AIDR. Measurements of HU and noise values were performed on images of the “HU linearity” module of the TOS-phantom. The module had five embedded objects, i.e., air, polypropylene, nylon, acrylic, and Delrin. On each object, a circle area of 4.32 cm2 was drawn and used to measure HU and noise values. The R2 of the relation between mass densities vs. HU values was used to measure HU linearities at four different tube voltages. The Mann-Whitney U test was used to compare unpaired data and p-value < 0.05 was considered statistically significant. Results: The AIDR method produced a significant smaller image noise than the FBP method (p-value < 0.05). There were no significant differences in HU values of images reconstructed using FBP and AIDR methods (p-value > 0.05). The HU values acquired by the methods showed the same linearity marked by coinciding linear lines with the same R2 value (> 0.999). Conclusion: AIDR methods produce the HU linearity as FBP methods with a smaller image noise level.

Download for Free

Sign up for free to access the full text

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

منابع من

با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

ذخیره در منابع من قبلا به منابع من ذحیره شده

{@ msg_add @}


عنوان ژورنال

دوره 10  شماره 2

صفحات  215- 224

تاریخ انتشار 2020-04-01

با دنبال کردن یک ژورنال هنگامی که شماره جدید این ژورنال منتشر می شود به شما از طریق ایمیل اطلاع داده می شود.

میزبانی شده توسط پلتفرم ابری doprax.com

copyright © 2015-2023