Uncertainties in effective dose estimation for CT transmission scan in total body PET-CT imaging with Auto mA3D tube current modulation

Authors

  • M.R. Ay Research Center for Molecular and Cellular Imaging, and Department of Medical Physics and Biomedical Engineering, Tehran University of Medical Sciences, Tehran, Iran
  • M.R. Deevband Department of Medical Engineering and Physics, Shahid Beheshti University of Medical Sciences, Tehran, Iran
  • P. Ghafarian hronic Respiratory Diseases Research Center, National Research Institute of Tuberculosis and Lung Diseases (NRITLD), Shahid Beheshti University of Medical Sciences, Tehran, Iran
  • Y. Salimi Department of Medical Engineering and Physics, Shahid Beheshti University of Medical Sciences, Tehran, Iran
Abstract:

Background: Positron Emission Tomography-Computed Tomography (PET-CT) is a useful hybrid imaging modality in the diagnosis of various malignancies. This modality imposes almost 20 mSv radiation dose to the patient. The purpose of the present study was to evaluate the uncertainties in calculated CT effective dose in TUBE CURRENT MODULATION-activated scans by Impact-Dose software.  Materials and Methods: Sixty total body DICOM (30 male and 30 female) whole body PET-CT images were selected. Volume CT Dose Index (CTDIvol) was recorded for each of the procedures. The image was divided into 5 regions of head & neck, chest, abdomen, pelvis and lower limbs according to special anatomical markers. Effective doses for total body and separate organs were calculated by means of Impact-Dose software once with global CTDIvol and once with a summation of doses calculated by 5 Regional CTDIvol and related scan ranges. Results: The difference among effective doses for some organs and total body were considerable. The mean and standard deviation (SD) of the coefficient of variations (CV%) for total body, breast, gonads, liver, lung, red bone marrow (RBM), thyroid, kidneys, and uterus were 12.56, 11.61, 9.44, 8.1, 11.31, 5.93, 8.61, 6.03 and 12.49, respectively. Uncertainties were higher for smaller patients by 19 noise indexes while these changes were higher for bigger patients and 22 noise indexes. Conclusion: The tube current variation depends on the acquisition and patient parameters. For measuring and reporting the total body and organs’ effective doses in order to estimate the risks of CT’s radiation for total body PET-CT procedures, the tube current variations must be considered.

Upgrade to premium to download articles

Sign up to access the full text

Already have an account?login

similar resources

Estimation of fetal absorbed dose from low-dose attenuation-correction CT in PET / CT imaging by using the Body Builder Phantom

One of the methods of studying the physiology and metabolism of important tissues such as the heart, brain and cancer tumors is the use of PET/CT System. A small number of patients are pregnant women who undergo a PET/CT scan due to lack of knowledge about pregnancy or due to dire clinical need. The final dose received by the fetus is based on three factors: the absorbed dose of the fetal tissu...

full text

Estimate bowtie filter shape in PET/CT scan with TLD

Introduction: The CT machine utilizes a bowtie filter to shape the X-ray beam and remove lower energy photons. Configuration of this bowtie filter is complex and its geometry is often not available in detail. It causes the CT dose index (CTDI) be with the different values in measurement versus Monte Carlo simulation studies and other analytical calculations. It is important esp...

full text

stimate bowtie filter shape in PET/CT scan with TLD

Introduction: The CT machine utilizes a bowtie filter to shape the X-ray beam and remove lower energy photons. Configuration of this bowtie filter is complex and its geometry is often not available in detail. It causes the CT dose index (CTDI) be with the different values in measurement versus Monte Carlo simulation studies and other analytical calculations. It is important esp...

full text

Effective dose estimation for oncological and neurological PET/CT procedures

BACKGROUND The aim of this study was to retrospectively evaluate the patient effective dose (ED) for different PET/CT procedures performed with a variety of PET radiopharmaceutical compounds. PET/CT studies of 210 patients were reviewed including Torso (n = 123), Whole body (WB) (n = 36), Head and Neck Tumor (HNT) (n = 10), and Brain (n = 41) protocols with 18FDG (n = 170), 11C-CHOL (n = 10), 1...

full text

False-positive FDG PET CT Scan in Vertebral Hemangioma

FDG PET CT scan is considered to be a sensitive tool to detect skeletal metastasis in known malignancies. However, it’s high sensitivity and low specificity may account for false positive diagnosis in cases of trauma, infection, inflammation and other benign conditions. Skeletal hemangioma is one of the common benign conditions which are typically ametabolic on FDG PET CT with no uptake on bone...

full text

My Resources

Save resource for easier access later

Save to my library Already added to my library

{@ msg_add @}


Journal title

volume 16  issue 4

pages  465- 472

publication date 2018-10

By following a journal you will be notified via email when a new issue of this journal is published.