Quantification and modelling of the dosimetric impact of the treatment couch in volumetric modulated arc therapy (VMAT)

Authors

  • J. Fleckenstein Department of Radiation Oncology, Mannheim Medical Center, Germany
  • M.C. Miao Department of Radiotherapy, The Fourth Hospital of Hebei Medical University, Shijiazhuang, China
  • R.H. Zhang Department of Biomedical Engineering,Tianjin University, Tianjin, China
  • W.W. Bai Department of Radiotherapy, The Fourth Hospital of Hebei Medical University, Shijiazhuang, China
  • Y.L. Gao The Physical Examination Center, Hebei General Hospital, Shijiazhuang, China
  • Z.F. Chi Department of Radiotherapy, The Fourth Hospital of Hebei Medical University, Shijiazhuang, China
Abstract:

Background: As the volumetric modulated arc therapy (VMAT) becoming a main role of treatment ways, the effect of couch top becomes more significant. It is imperative to re-evaluate the couches that previously may have been considered of no importance during early treatment techniques. The impact of couch top on radiation delivery was explored and the couch model was tested with the aim of reducing the couch absorption influences.  Materials and Methods: Attenuation measurements were performed in a cylindrical phantom with an ionization chamber positioned at the isocenter. Couch model was obtained by importing its actual CT scan, and the accuracy was evaluated by comparing percentage deviation at 2 and 5 mm voxel grid size. Effects on surface dose were measured using EBT3 film with the constant SSD at different depths and beam energies at the gantry angle 180° and 0°, respectively. Results: Couch top increases surface dose from 45.9 % to 95.8 % , from 35.0 % to 87.9 % and from 29.2% to 73.9 % for 10 cm ×10 cm field at 6 ,10 and 18 MV, respectively. Due to the couch absorption the case of vertebral metastasis VMAT plan D50 of the PTV changed from 30 Gy to 29.3 Gy. Couch model with uniform electron density of 0.18g/cm3 demonstrated an excellent agreement between measured and TPS computed dose. Conclusion: The treatment couch presence between the patient and beam source significantly alters dose in the patient. Modelling the couch in the Monaco TPS can adequately predict the altered dose distribution.  

similar resources

Dosimetric Evaluation of Volumetric Modulated Arc Therapy (VMAT) and Intensity Modulated Radiotherapy (IMRT) Using AAPM TG 119 Protocol

Background: The commissioning accuracy of Volumetric Modulated Arc Therapy (VMAT) need to be evaluated.Objective: To test and evaluate commissioning accuracy of VMAT based on the TG 119 protocols at local institution. Material and Methods: The phantom, structure sets, VMAT and IMRT beam parameter setup, dose prescriptions and planning objectives were following TG 119 guidelines to c...

full text

Volumetric modulated arc therapy ( VMAT )

New calorimeters Simon Duane, NPL Calorimetry remains the most fundamental method for the measurement of absorbed dose and is the basis of almost all absorbed dose primary standards worldwide. Absorbed dose to water is realised either directly, by using water as the absorbing medium, or indirectly, by using another medium (usually graphite) and determining a factor to convert dose from that med...

full text

Commissioning of volumetric modulated arc therapy (VMAT).

PURPOSE Volumetric modulated arc therapy (VMAT) involves the simultaneous use of dynamic multileaf collimator (DMLC) techniques and gantry arcing; appropriate quality assurance is therefore required. This article describes the development and implementation of procedures for commissioning VMAT on a commercial linear accelerator (Elekta PreciseBeam VMAT with MLCi and Beam Modulator heads). MAT...

full text

Quantification and comparison the dosimetric impact of two treatment couch model in VMAT

The use of Monte Carlo treatment planning systems (TPS) in radiation therapy has increased the dosimetric accuracy of VMAT treatment sequences. However, this accuracy is compromised by not including the treatment couch into the treatment planning process. Therefore, the impact of the treatment couch on radiation delivery output was determined, and two different couch models (uniform couch model...

full text

Ultrafast treatment plan optimization for volumetric modulated arc therapy (VMAT).

PURPOSE To develop a novel aperture-based algorithm for volumetric modulated are therapy (VMAT) treatment plan optimization with high quality and high efficiency. METHODS The VMAT optimization problem is formulated as a large-scale convex programming problem solved by a column generation approach. The authors consider a cost function consisting two terms, the first enforcing a desired dose di...

full text

Comparison of hybrid volumetric modulated arc therapy (VMAT) technique and double arc VMAT technique in the treatment of prostate cancer

BACKGROUND Volumetric modulated arc therapy (VMAT) has quickly become accepted as standard of care for the treatment of prostate cancer based on studies showing it is able to provide faster delivery with adequate target coverage and reduced monitor units while maintaining organ at risk (OAR) sparing. This study aims to demonstrate the potential to increase dose conformality with increased plann...

full text

My Resources

Save resource for easier access later

Save to my library Already added to my library

{@ msg_add @}


Journal title

volume 17  issue 2

pages  335- 344

publication date 2019-04

By following a journal you will be notified via email when a new issue of this journal is published.

Keywords

Hosted on Doprax cloud platform doprax.com

copyright © 2015-2023