Proof of Symmetric Dose Distribution in Mammosite Applicator in Breast Brachytherapy with MCNP Simulation

Authors

  • Alireza Shirazi Hosseini Dokht Professor of Physics and Medical Engineering Department of University of Tehran
  • Fatemeh Mahdavi M.Sc. of Medical Radiation Engineering, of Science and Research University of Tehran
  • Mahdi Sadeghi Professor of Medical Physics Department of Medical University of Iran
Abstract:

Introduction: In brachytherapy treatments, high-dose sources are used in interstitial placement. In this paper, the modeling of a common breast cancer treatment applicator called MammoSite is done by the Monte Carlo simulation code. Then, to study and calculate the amount of doses in the breast and the dose rate of the organs at risk of radiation, including: lungs, ribs and skin, Then the absorbed dose ratio is calculated in the target volume to the treated chest. Materials and Methods: The MCNPX code: is the based on the Monte Carlo method. The code, with input file information and the use of the cross-section library, solves the problem and produces results in an output file. MIRD phantom: To calculate the dose, a model of the human body is required as a phantom. All organs of the human body are characterized by a detailed description for the calculation of the absorbed dose in this phantom. Mammosite applicator: This device has a catheter that can be attached to the high dose rate after loading machine that connect to the source. The end of catheter is surrounded by a balloon. The most common size of this balloon is 4-5 cm and 5-6 cm, which has a volume between 34-113 cm cubic meters. Results: As expected, the absorbed dose in the target volume and the treated breast was the highest and after that the lung received the same significant dose. That is, the highest doses of the target or tumor volume and the lower doses of the surrounding organs are reached. Healthy organs have received around high dose ratios. When the right breast was treated, the right lungs, ribs, and skin had the highest dose ratios relative to the dose received by the tumor and when the left breast was treated, the left lung, ribs and skin had the highest dose ratios relative to the dose received by the tumor. Conclusion: Given that in Iran, not only the mammosite applicator but also the treatment planning system for brachytherapy is not available. The inability of the MammoSite applicator in the asymmetric dose distribution cannot be shown. However, due to the absorption dose ratio of breast tissue to absorbed dose by the target volume in the right breast, which is numerically equal to 1.035, And the absorbed dose rate of the left breast tissue to the absorbed dose by the target volume, which is a number equal to 1.033, both of which are approximately equal, It can be concluded that the same dose that reached the tumor tissue is about the same size as the healthy tissue of the treated breast. This represents a symmetric dose and proves that the MammoSite applicator is incapable of asymmetric dispensing.

Upgrade to premium to download articles

Sign up to access the full text

Already have an account?login

similar resources

Calculation of absorbed dose rate in the lungs, ribs and skin by the mammosite applicator in breast cancer brachytherapy with MCNP

Introduction: Breast cancer is the most common cancer among women. Brachytherapy is one of the ways to treat breast cancer. In brachytherapy treatments, high-dose sources are used in interstitial placement. Here, Iridium-192 is used. This source emits gamma rays with energies ranging from 136 kV to 1060 kV electron volts. The activity of the used spring is 10 Ci. The simulated ...

full text

Consideration of accurate dose distribution for Contura balloon catheter in breast brachytherapy with MCNP5 simulation

Introduction: Contura is a balloon applicator that used to deliver intra cavity radiation for Partial Breast Irradiation (PBI) treatments. Contura with 4 peripheral lumens surrounding the central, for loading HDR sources from an after loader system can provide an asymmetric dose distribution and gives a more flexible shape of dose than earlier balloon applicators such as Mammo...

full text

The impact of uncertainties associated with MammoSite brachytherapy on the dose distribution in the breast

The MammoSite radiation therapy system is a novel technique for treatment of patients with early-stage breast cancer. It was developed to overcome the longer schedules associated with external-beam radiation therapy. It consists of a small balloon (4 cm in diameter) connected to an inflation channel and a catheter for the passage of a high dose rate 192Ir brachytherapy source. The device is pla...

full text

A Comparison of Skin Dose Delivered with MammoSite and Multicatheter Breast Brachytherapy

Background: Accelerated partial breast irradiation via interstitial balloon brachytherapy is a fast and effective treatment method for certain early stage breast cancers however skin, chest wall and Lung doses are correlated with toxicity in patients treated with breast brachytherapy.Objective: To investigate the percentage of the dose received by critical organ (skin), thermoluminescence detec...

full text

A Comparison of Skin Dose Delivered with MammoSite and Multicatheter Breast Brachytherapy

BACKGROUND Accelerated partial breast irradiation via interstitial balloon brachytherapy is a fast and effective treatment method for certain early stage breast cancers however skin, chest wall and Lung doses are correlated with toxicity in patients treated with breast brachytherapy. OBJECTIVE To investigate the percentage of the dose received by critical organ (skin), thermoluminescence dete...

full text

Skin and chest wall dose with multi-catheter and MammoSite breast brachytherapy: Implications for late toxicity.

PURPOSE Accelerated partial breast irradiation (APBI) continues to increase in popularity. Up to 14% of patients treated with the MammoSite (MS) report some degree of chronic pain, which may be related to chest wall toxicity. Reports from several institutions using the multicatheter (MC) technique have not shown associated elevated chest wall toxicity. Additionally, a recent investigation has s...

full text

My Resources

Save resource for easier access later

Save to my library Already added to my library

{@ msg_add @}


Journal title

volume 15  issue Special Issue-12th. Iranian Congress of Medical Physics

pages  232- 232

publication date 2018-12-01

By following a journal you will be notified via email when a new issue of this journal is published.

Hosted on Doprax cloud platform doprax.com

copyright © 2015-2023