Measuring midline dose without build-up cap for patients with brain tumor undergoing 15MV external radiotherapy by using EBT3 Gafchromic film
Authors
Abstract:
Introduction: The Purpose of this study has been measuring midline dose for the patients undergoing brain radiotherapy without any need to build-up cap. Materials and Methods: The study was performed on 28 patients with brain tumor undergoing 15 MV radiotherapy by a Siemens LINAC. To estimate the midline dose, the entrance and exit doses were calculated by a conversion ratio. The conversion ratio was determined as the ratio of the measured dose by EBT3 Gafchromic film at the phantom surface to that measured by an ionization chamber at the depth of 3 and 12 cm. Then, the midline and exit transmission curves were obtained in a solid water phantom for various irradiation conditions with and without a 30-degree wedge for various field sizes of 5×5, 10×10 and 15×15 cm2 and at various SSDs including 80, 90 and 100 cm at various depths ranged from 6 to 22 cm by a 0.6 CC ionization chamber. Eventually, to measure the midline dose, we used both of the midline transmission and arithmetic mean algorithms. Results: The mean and standard deviation of the whole sample for midline transmission algorithm were -3% and 4.52% and for that of arithmetic mean algorithm -2.97% and 4.3%, respectively. However, for 9 patients, more than 5% error was noted in estimated midline doses. Conclusion: A new method was developed for measuring the midline dose without any need to use the build-up cap. The main advantage of the developed method is that there is no disruption in the dose reached to the treatment volume.
similar resources
دزیمتری in-vivo در پرتودرمانی خارجی تومورهای مغزی تحت درمان با فوتون MV15 با استفاده از فیلم گاف کرومیک EBT3
Background and purpose: In-vivo dosimetry is used to ensure accurate delivery of dose to tumors during radiotherapy. This study aimed at measuring the entrance doses without the build-up cap for patients undergoing 15MV radiotherapy and determining the percentage errors between doses measured by films and those calculated with a treatment planning system (TPS). Materials and methods: The stu...
full textPrecise film dosimetry for stereotactic radiosurgery and stereotactic body radiotherapy quality assurance using Gafchromic™ EBT3 films
PURPOSE The purpose of this study is to evaluate the dosimetric uncertainty associated with Gafchromic™ (EBT3) films and establish a practical and efficient film dosimetry protocol for Stereotactic Radiosurgery (SRS) and Stereotactic Body Radiotherapy (SBRT). METHOD AND MATERIALS EBT3 films were irradiated at each of seven different dose levels between 1 and 15 Gy with open fields and standar...
full textUnlaminated Gafchromic Ebt3 Film for Ultraviolet Radiation Monitoring.
Measurement of ultraviolet (UV) radiation is important for human health, especially with the expanded usage of short wavelength UV for sterilization purposes. This work examines unlaminated Gafchromic EBT3 film for UV radiation monitoring. The authors exposed the film to select wavelengths in the UV spectrum, ranging from 207 to 328 nm, and measured the change in optical density. The response o...
full textDosimetric characterization and use of GAFCHROMIC EBT3 film for IMRT dose verification
Radiochromic film has become an important tool to verify dose distributions in highly conformal radiation therapy such as IMRT. Recently, a new generation of these films, EBT3, has become available. EBT3 has the same composition and thickness of the sensitive layer of the previous EBT2 films, but its symmetric layer configuration allows the user to eliminate side orientation dependence, which i...
full textDetermining rectal dose through cervix cancer radiotherapy by 9 MV photon beam using TLD and XR type T GAFCHROMIC® Film
Background: The goals of the present research were to investigate the rectal dose during four chosen techniques of cervical cancer radiotherapy and to examine how accurately the treatment planning represents dose measurements, and it’s practicality for routine use as well as, to determine the homogeneity of dose in tumor volume. Materials and Methods: The study was carried out using a Nepton 10...
full textEvaluation of lens dose from anterior electron beams: comparison of Pinnacle and Gafchromic EBT3 film
Lens dose is a concern during the treatment of facial lesions with anterior electron beams. Lead shielding is routinely employed to reduce lens dose and minimize late complications. The purpose of this work is twofold: 1) to measure dose pro-files under large-area lead shielding at the lens depth for clinical electron energies via film dosimetry; and 2) to assess the accuracy of the Pinnacle tr...
full textMy Resources
Journal title
volume 15 issue Special Issue-12th. Iranian Congress of Medical Physics
pages 63- 63
publication date 2018-12-01
By following a journal you will be notified via email when a new issue of this journal is published.
Hosted on Doprax cloud platform doprax.com
copyright © 2015-2023