Manual production and quality control of 68Ga-PSMA-11 radiopharmaceutical for PET-CT imaging of prostate cancer: optimization and 60 production series experience
Authors
Abstract:
Background: 68Ga-PSMA-11, a novel PET radiopharmaceutical, is currently used for imaging of almost all prostate cancers. However, automatic radiolabeling procedures have been recommended. In this research, according to the chemical nature of the PSMA-11, manual, optimized and safe production of this radiopharmaceutical has been investigated and a standard method has been presented to achieve a repeatable quality with the use of an Iran made 68Ge/68Ga generator. Methods: The effect of the PSMA-11 weight, temperature, pH, and radiolabeling time on the radiochemical purity (RCP) was evaluated. The effectiveness of the optimized condition on the quality of radiopharmaceutical in 60 production series was studied according to US Pharmacopeia guidelines. Also, by monitoring the dose rate of the personnel, the safety of the manual production was checked. Results: Increasing the temperature from room temperature (RT) to 40◦C enhanced the RCP of 10 μg of PSMA-11 to acceptable ranges of > 90%. The weights of 25 and 50 μg had RCPs>90% at RT. The appropriate pH range was determined to be 5-6. The 10 minutes’ reaction time yielded a repeatable RCP>90% for the 10 μg PSMA-11 in 60 production series. Body and hands dose rates were 15% and 2.4% of the recommended annual dose rate limit for body and hands, respectively. Conclusion: The results indicate the high quality of the Iran made 68Ge/68Ga generator. PET-CT centers can in-site, manually, safely and with repeatable quality produce 68Ga-PSMA-11 for prostate cancer imaging away from the challenge of transport problems.
similar resources
Direct comparison of 99mTc-PSMA SPECT/CT and 68Ga-PSMA PET/CT in patients with prostate cancer
Objective(s): 99mTc-PSMA SPECT/CT is a cost effective alternative for 68Ga-PSMA PET/CT. The aim of this study was to directly compare these two techniques in patients with prostate cancer. Methods: 28 man with prostate cancer were studied using 99mTc-PSMA SPECT/CT and 68Ga-PSMA PET/CT in a shor...
full text[68Ga]Ga-PSMA-11 PET/CT for staging and patient management of high-risk prostate cancer: A single-center experience from Iran
Introduction: Accurate staging plays an important role in management of patients with prostate cancer especially in high-risk group. Today, [68Ga]Ga-PSMA-11 PET/CT should be considered as the preferred imaging tool for treatment planning and initial staging of the disease. Methods: A total number of 628 patients with prostate cancer referred to...
full text68Ga-PSMA-11 PET/CT for prostate cancer staging and risk stratification in Chinese patients
We evaluated the clinical utility of 68Ga-PSMA-11 PET/CT for staging and risk stratification of treatment-naïve prostate cancer (PCa) and metastatic castrate-resistant prostate cancer (mCRPC). Twenty-two consecutive patients with treatment-naïve PCa and 18 with mCRPC were enrolled. 68Ga-PSMA-11 PET/CT and magnetic resonance imaging (MRI) were performed for the evaluation of primary prostatic le...
full textNovel Preclinical and Radiopharmaceutical Aspects of [68Ga]Ga-PSMA-HBED-CC: A New PET Tracer for Imaging of Prostate Cancer
The detection of prostate cancer lesions by PET imaging of the prostate-specific membrane antigen (PSMA) has gained highest clinical impact during the last years. 68Ga-labelled Glu-urea-Lys(Ahx)-HBED-CC ([68Ga]Ga-PSMA-HBED-CC) represents a successful novel PSMA inhibitor radiotracer which has recently demonstrated its suitability in individual first-in-man studies. The radiometal chelator HBED-...
full text68Ga-PSMA PET-CT Imaging of Metastatic Adenoid Cystic Carcinoma
A patient with a history of adenoid cystic carcinoma of the nasal cavity presented himself with bone pain and an elevated PSA level. On suspicion of metastatic prostate cancer a 68Ga-PSMA PET-CT was performed. The PET-CT showed numerous lung and non-sclerotic bone metastasis. Biopsy of a bone metastasis was performed and pathology showed adenoid cystic carcinoma instead of prostate cancer. Immu...
full textOptimized production and quality control of 68Ga-DOTATATE
Introduction: Optimized production and quality control of 68Ga-DOTATATE as an efficient and preferable PET radiotracer for somatostatin receptor imaging in neuroendocrine tumors is of great interest. In this study effort has been made to present a fast, efficient, cost-effective and facile protocol for 68Ga-DOTATATE productions for clinical trials. Methods:68Ga...
full textMy Resources
Journal title
volume 26 issue 4
pages 32- 43
publication date 2019-07
By following a journal you will be notified via email when a new issue of this journal is published.