Edge Artifacts in Point Spread Function-based PET Reconstruction in Relation to Object Size and Reconstruction Parameters

Authors

  • Kazuhiko Himuro Division of Radiology, Department of Medical Technology, Kyushu University Hospital, Fukuoka, Japan
  • Masayuki Sasaki Department of Health Sciences, Faculty of Medical Sciences, Kyushu University, Fukuoka, Japan
  • Shingo Baba Department of Clinical Radiology, Kyushu University Hospital, Fukuoka, Japan
  • Shinichi Awamoto Division of Radiology, Department of Medical Technology, Kyushu University Hospital, Fukuoka, Japan
  • Yoshiyuki Umezu Division of Radiology, Department of Medical Technology, Kyushu University Hospital, Fukuoka, Japan
  • Yuji Tsutsui Division of Radiology, Department of Medical Technology, Kyushu University Hospital, Fukuoka, Japan
Abstract:

Objective(s): We evaluated edge artifacts in relation to phantom diameter and reconstruction parameters in point spread function (PSF)-based positron emission tomography (PET) image reconstruction.Methods: PET data were acquired from an original cone-shaped phantom filled with 18F solution (21.9 kBq/mL) for 10 min using a Biograph mCT scanner. The images were reconstructed using the baseline ordered subsets expectation maximization (OSEM) algorithm and the OSEM with PSF correction model. The reconstruction parameters included a pixel size of 1.0, 2.0, or 3.0 mm, 1-12 iterations, 24 subsets, and a full width at half maximum (FWHM) of the post-filter Gaussian filter of 1.0, 2.0, or 3.0 mm. We compared both the maximum recovery coefficient (RCmax) and the mean recovery coefficient (RCmean) in the phantom at different diameters.Results: The OSEM images had no edge artifacts, but the OSEM with PSF images had a dense edge delineating the hot phantom at diameters 10 mm or more and a dense spot at the center at diameters of 8 mm or less. The dense edge was clearly observed on images with a small pixel size, a Gaussian filter with a small FWHM, and a high number of iterations. At a phantom diameter of 6-7 mm, the RCmax for the OSEM and OSEM with PSF images was 60% and 140%, respectively (pixel size: 1.0 mm; FWHM of the Gaussian filter: 2.0 mm; iterations: 2). The RCmean of the OSEM with PSF images did not exceed 100%.Conclusion: PSF-based image reconstruction resulted in edge artifacts, the degree of which depends on the pixel size, number of iterations, FWHM of the Gaussian filter, and object size.

Upgrade to premium to download articles

Sign up to access the full text

Already have an account?login

similar resources

Point spread function based image reconstruction in optical projection tomography.

As a result of the shallow depth of focus of the optical imaging system, the use of standard filtered back projection in optical projection tomography causes space-variant tangential blurring that increases with the distance to the rotation axis. We present a novel optical tomographic image reconstruction technique that incorporates the point spread function of the imaging lens in an iterative ...

full text

Improvement in PET/CT image quality with a combination of point-spread function and time-of-flight in relation to reconstruction parameters.

UNLABELLED The aim of this study was to investigate the effects of the point-spread function (PSF) and time-of-flight (TOF) on improving (18)F-FDG PET/CT images in relation to reconstruction parameters and noise-equivalent counts (NEC). METHODS This study consisted of a phantom study and a retrospective analysis of 39 consecutive patients who underwent clinical (18)F-FDG PET/CT. The body phan...

full text

Reconstruction vs. Interaction-based Output Practice: (in relation to EFL learner’s speaking skill and learning styles)

  The belief that output practice is crucial in L2 learning affects foreign language teaching methodology. And researchers have endeavored to find the best ways to encourage learners to produce and practice whatever they hear as an input in the process of learning. Moreover, learning styles and the importance of matching learners’ styles with those of teachers inspired the researchers to inves...

full text

Evaluation of the Effect of Tumor Position on Standardized Uptake Value Using Time-of-Flight Reconstruction and Point Spread Function

Objective(s): The present study was conducted to examine whether the standardized uptake value (SUV) may be affected by the spatial position of a lesion in the radial direction on positron emission tomography (PET) images, obtained via two methods based on time-of-flight (TOF) reconstruction and point spread function (PSF). Methods: A cylinder phantom with the sphere (30mm diameter), located in...

full text

Evaluation of spatial dependence of point spread function-based PET reconstruction using a traceable point-like 22Na source

BACKGROUND The point spread function (PSF) of positron emission tomography (PET) depends on the position across the field of view (FOV). Reconstruction based on PSF improves spatial resolution and quantitative accuracy. The present study aimed to quantify the effects of PSF correction as a function of the position of a traceable point-like 22Na source over the FOV on two PET scanners with a dif...

full text

My Resources

Save resource for easier access later

Save to my library Already added to my library

{@ msg_add @}


Journal title

volume 5  issue 2

pages  141- 151

publication date 2017-06-01

By following a journal you will be notified via email when a new issue of this journal is published.

Hosted on Doprax cloud platform doprax.com

copyright © 2015-2023