Dosimetric influence of Flattening Filter (FF) and Flattening Filter Free (FFF) 6 and 10 MV photon beams on Volumetric Modulated Arc Therapy (VMAT) planning in case of prostate carcinoma
Authors
Abstract:
Background: In the treatment of prostate cancer, radiotherapy is the potential to increase second primary cancers such as bladder and rectal cancers. The reasons for this potential are more monitor units (MUs), therefore a larger total body dose because of leakage radiation, a bigger volume of normal tissue is exposed to lower radiation doses. This study was designed to compare the integral dose of using Flattening Filter (FF) and Flattening Filter-Free (FFF) 6 and 10 MV photon beams via volumetric modulated arc therapy (VMAT) for prostate cancer patients. Materials and Methods: Twenty prostate cancer patients were selected retrospectively for this planning study. VMAT plans were developed using the Eclipse (Varian Medical System, Palo Alto, California, USA) Treatment Planning System (TPS) with 6 MV FF, 6 MV FFF, 10 MV FF and 10 MV FFF for each patient. Conformity index (CI), homogeneity index (HI), Integral dose (ID), the volume receiving 5 Gy (V5%) and monitor units (MUs) were compared. Results: The use of 10 MV FF had 206 liter*Gy integral dose to Body-CTV volume. Using 10 MV FFF had 204 liter*Gy integral dose to normal structures. When 10 MV FF or 10 MV FFF were used instead of 6 MV FF and 6 MV FFF integral dose decreased as -7% and -8%, respectively. The dosimetric difference were statistically significant (p<0.05). The use of 10 MV FFF rather than 10 MV FF had limited influence on the integral dose for rectum, bladder, penile bulb and femoral heads. Conclusion: This study showed that high energy photons (10 MV FF, 10 MV FFF) have lower integral dose than low energy photons (6 MV FF, 6 MV FFF). The relationship between low energies, high energies and integral doses is significant, although there is no significant relationship between V5% doses of all energies. In comparison to different treatment plans, we showed that V5% alone did not provide enough information when possible secondary cancer risks were calculated.
similar resources
Investigation of Neutron Contamination of Flattening Filter and Flattening Filter-Free 10-MV Photon Beams in Elekta InfinityTM Accelerator
Introduction: This study aimed to measure the neutron contamination of flattening filter (FF) and flattening filter-free (FFF) 10-MV photon beams delivered by the Elekta InfinityTM accelerator. Material and Methods: The photoneutron spectrum produced by the Linac head was evaluated using a Monte Carlo (MC) simulation. The geometry ...
full textSurface dose variations in 6 and 10 MV flattened and flattening filter‐free (FFF) photon beams
As the use of linear accelerators operating in flattening filter-free (FFF) modes becomes more widespread, it is important to have an understanding of the surface doses delivered to patients with these beams. Flattening filter removal alters the beam quality and relative contributions of low-energy X-rays and contamination electrons in the beam. Having dosimetric data to describe the surface do...
full textComparison of VMAT‐SABR treatment plans with flattening filter (FF) and flattening filter‐free (FFF) beam for localized prostate cancer
The purpose of this study is to investigate the feasibility of using a flattening filter-free (FFF) beam with an endorectal balloon for stereotactic ablative body radiotherapy (SABR) of clinically localized prostate cancer. We assessed plans of SABR with volumetric-modulated arc therapy (VMAT) that used a flattening filter (FF) beam and an FFF beam and compared the verification results of dosim...
full textDosimetric Validation of Volumetric Modulated Arc Therapy (VMAT) Using AAPM TG-119 Benchmark Plans in an Upgraded CLINAC 2100CD for Flattening Filter Free (FFF) Photon Beams
Background: Recently we have upgraded our Varian Clinac 2100CD with a 6MV FFF beam, this upgrade being the first of its kind in our country. Even though the dosimetric characteristics of FFF beams have been reported both in experimental and Monte Carlo studies, application in planning and delivery is complex. The aim of this study was to validate the commissioning of upgraded FFF beams dosimetr...
full textFlattening Filter-Free Beams in Intensity-Modulated Radiotherapy and Volumetric Modulated Arc Therapy for Sinonasal Cancer
PURPOSE To evaluate the dosimetric impacts of flattening filter-free (FFF) beams in intensity-modulated radiotherapy (IMRT) and volumetric modulated arc therapy (VMAT) for sinonasal cancer. METHODS For fourteen cases, IMRT and VMAT planning was performed using 6-MV photon beams with both conventional flattened and FFF modes. The four types of plans were compared in terms of target dose homoge...
full textPlanning Study of Flattening Filter Free Beams for Volumetric Modulated Arc Therapy in Squamous Cell Carcinoma of the Scalp
PURPOSE Flattening filter free (FFF) beams show the potential for a higher dose rate and lower peripheral dose. We investigated the planning study of FFF beams with their role for volumetric modulated arc therapy (VMAT) in squamous cell carcinoma of the scalp. METHODS AND MATERIALS One patient with squamous cell carcinoma which had involvement of entire scalp was subjected to VMAT using TrueB...
full textMy Resources
Journal title
volume 17 issue 2
pages 253- 263
publication date 2019-04
By following a journal you will be notified via email when a new issue of this journal is published.