Diagnostic Reference Level of Computed Tomography Examinations and Need for Dose Optimization in Ondo State, Nigeria
Authors
Abstract:
Introduction: The present study was conducted to obtain State diagnostic reference levels (DRLs) of five routine computed tomography (CT) examinations from two CT centers in Ondo State and to identify factors responsible for dose variation and escalation in these CT centers. Material and Methods: Acquisition parameters and CT dose indices were collected from the storage drives of the two CT centers namely Federal Medical Centre, Owo and Trauma Center, Ondo, Ondo State, Nigeria, for six months on electronic spreadsheets for cranial, sinus, chest, abdomen and pelvis examinations. In addition, dose indices for multiphase examinations were collected to analyze chest and abdominal doses. Wilcoxon rank-sum test was used to assess variations in dose distributions of the two health institutions. Results: The following diagnostic reference levels (DRLs) were obtained at 91 mGy; 1943 mGy.cm, 69 mGy; 1159 mGy.cm, 45 mGy; 1064 mGy.cm, 50 mGy; 2545 mGy.cm and 26 mGy; 622 mGy.cm in cranial, sinus, chest, abdomen and pelvis examinations respectively. Conclusion: Estimated State DRLs exceed national and other DRLs indicating that there is a need to improve the quality of CT-examination for a better benefit to risk ratio.However, benchmarking DRLs to median dose levels (Achievable dose levels) instead of the upper quartile will be a good starting point in achieving the optimal dose level.
similar resources
Patient-Specific Radiation Dose and Cancer Risk in Computed Tomography Examinations in Ondo, Nigeria
Introduction: The dose in computed tomography (CT) often approach or exceed the optimum levels, thereby increasing the probability of cancer induction. With wide application of this diagnostic test, it is expedient to determine the effective dose (ED) for each patient to estimate their cancer risk. This study was conducted to investigate the patient-specific dose (PSD) and cancer risk in CT exa...
full textDiagnostic Reference Levels for Computed Tomography Examinations in Iran: A Nationwide Radiation Dose Survey
Introduction: International Commission on Radiological Protection introduced three basic principles of radiation protection, namely justification, optimization, and dose limit. Medical exposure has no dose limits, and generally, diagnostic reference levels are used as a tool for optimization of patient protection. Material and Methods: Dosimetry was performed on 20 CT scanners located in 14 cit...
full textDose Assessment in Computed Tomography Examination and Establishment of Local Diagnostic Reference Levels in Mazandaran, Iran
Background: Medical X-rays are the largest man-made source of public exposure to ionizing radiation. While the benefits of Computed Tomography (CT) are well known in accurate diagnosis, those benefits are not risk-free. CT is a device with higher patient dose in comparison with other conventional radiation procedures. Objective: This study is aimed at evaluating radiation dose to patients from ...
full textPaediatric Dose AssessmentToward Establishment of National Diagnostic Reference Levels for Digital Radiography Examinations in Iran
Introduction: Paediatric patients deserve special attention because of the higher radiation risks compared with adults. The main purpose of this study is to determine patient dose and diagnostic reference level for Paediatric through Entrance Surface Dose (ESD) calculations in common X-ray examination and to establish Paediatric national DRLs in digital radiography in Iran. <...
full textEstimation of diagnostic reference levels for children computed tomography: A study in Tehran, Iran
Background: The Diagnostic reference levels (DRLs) play a critical role in the optimization of radiation dose especially, in some conditions like pediatrics. They are useful indicators by which the radiologists can be aware of delivered excess radiation doses to the patients, and take corrective actions if necessary. In order to meet some requirements for establishing the national computed tomo...
full textEstablishment of institutional diagnostic reference level for computed tomography with automated dose‐tracking software
INTRODUCTION The aim of this study was to establish institutional diagnostic reference levels (DRLs) by summarising doses collected across the five computed tomography (CT) system in our institution. METHODS CT dose data of 15940 patients were collected retrospectively from May 2015 to October 2015 in five institutional scanners. The mean, 75th percentile and 90th percentile of the dose sprea...
full textMy Resources
Journal title
volume 17 issue 4
pages 266- 272
publication date 2020-07-01