Comparing X-ray dose reduction capability of silicon-bismuth micro- and nanocomposite shields using chest CT test
Authors
Abstract:
Comparison of radiation shields of different thicknesses and structures (especially different particle sizes of their constituents) is helpful to select an appropriate shield for patient protection in medical imaging tests particularly CT-scan. In this study micro- and nano-powders of bismuth were mixed with silicone matrix to make composite radiation shields for breast. The shields were provided in dimensions of 20×20 cm, thickness of 1 mm, and bismuth weight ratio of 10%. The chest CT test (TOSHIBA 16 multi-slice device) was performed on a female phantom with normal breast size in 100 kVp and 50 mA with slice thickness of 0.5 mm and pitch of 1 mm. To avoid image artifacts, a foam layer of 1 cm was wrapped around the phantom under the shields. Measurement of dose in the surface (equivalent to skin) and forth layer (equivalent to glandular tissue) of phantom breast was done using thermoluminescence dosimeter. Results showed that dose on the surface of phantom breast declined up to 12% and 18.4% in the presence of silicon-bismuth micro- and nanocomposite shields, respectively, and both composites reduced the breast dose significantly (p<0.5). Also, the nanocomposite shields reduced radiation dose more than the microcomposite shields. Additionally, less noise variation was observed in CT images acquired with the nanocomposite shields.
similar resources
Designing and construction of breast shields using silicone composite of Bismuth for chest CT
Background: Bismuth composites are new and non-lead shields that are capable of breast dose reduction while preserving image quality. In this study, different percentages of Bismuth were used in designing and constructing polymer composites as breast shields in chest CT. Materials and Methods: Micro particles of Bismuth with a ratio of 1% to 15% in Silicone matrix with two thicknesses were use...
full textBismuth-Silicon and Bismuth-Polyurethane Composite Shields for Breast Protection in Chest Computed Tomography Examinations
The article aims at constructing protective composite shields for breasts in chest computed tomography and investigating the effects of applying these new bismuth composites on dose and image quality. Polyurethane and silicon with 5% of bismuth were fabricated as a protective shield. At first, their efficiency in attenuating the X-ray beam was investigated by calculating the total attenuation c...
full textRadiation attenuation properties of shields containing micro and Nano WO3 in diagnostic X-ray energy range
Background: It has recently been shown that the particle size of materials used for radiation shielding can affect the magnitude of radiation attenuation. Over the past years, application of nano-structured materials in radiation shielding has attracted attention world-wide. The purpose of this study was to investigate the shielding properties of the lead-free shields containing micro and nano-...
full textradiation attenuation properties of shields containing micro and nano wo3 in diagnostic x-ray energy range
background: it has recently been shown that the particle size of materials used for radiation shielding can affect the magnitude of radiation attenuation. over the past years, application of nano-structured materials in radiation shielding has attracted attention world-wide. the purpose of this study was to investigate the shielding properties of the lead-free shields containing micro and nano-...
full textPatient effective dose evaluation for chest X-ray examination in three digital radiography centers
Background: The radiation doses resulting from diagnostic X-ray examinations are routinely measured in terms of entrance surface dose (ESD) and effective dose (ED). In this study, for the purpose of radiation protection, the radiation doses received from Digital chest X-ray examination were evaluated in terms of ESD and ED. Material and Methods: The ED was calculated by using the MCNP ...
full text[Breast dose reduction in female CT screening for lung cancer using various metallic shields].
We evaluated the effectiveness of metallic shields that were used for reduction of the breast dose in thoracic computed tomography(CT). For the evaluation, we measured breast surface dose and image standard deviation(SD)in the lung area. The metallic shields were made from bismuth, zinc, copper, and iron. The bismuth shield has been marketed and used for dose reduction. The other three metallic...
full textMy Resources
Journal title
volume 7 issue 3
pages 35- 40
publication date 2019-09