Comparative Study between Electronic Portal Imaging Device (EPID) and Cone Beam Computed Tomography (CBCT) for Radiation Treatment Verifications

Authors

  • Eyachew Misganew Tegaw PhD student in Medical Physics, Department of Medical Physics and Biomedical Engineering, School of Medicine, Tehran University of Medical Sciences, International Campus (TUMS-IC), Tehran, Iran
  • Getu Ferenji Tadesse PhD student in Medical Physics, Department of Medical Physics and Biomedical Engineering, School of Medicine, Tehran University of Medical Sciences, International Campus (TUMS-IC), Tehran, Iran
  • Ghazale Geraily Doctor of Medical Physics, Department of Medical Physics and Biomedical Engineering, Tehran University of Medical Sciences, Tehran, Iran
Abstract:

Introduction: Electronic Portal Imaging Device (EPID) and Cone Beam Computed Tomography (CBCT) are the preferred tools of Image Guided Radiotherapy (IGRT) and Dose Guided Radiotherapy (DGRT) which have been used for Radiotherapy treatment verifications. As a result, the number of publications dealing with these two tools for radiation treatment verification has increased considerably over the past years. This paper aims to compare and contrast the overall differences and similarities of EPID and CBCT tools based on their set-up verification accuracy, dosimetric verification accuracy, calibration and quality control methods, image quality, dose measurement, calculation, and evaluation methods for radiation treatment verifications.   Materials and Methods: The relevant articles published in English were searched based on the keywords of EPID and CBCT for radiation treatment verifications using PubMed, EMBASE, Scopus, and Web of Science databases and finally exported to EndNote for screening, downloading, and reference management processes.   Results: A total of 2512 articles were searched and subsequently screened by removing duplications and reading title and abstract to identify original articles that directly related with radiation treatment verifications using EPID and CBCT for different cases. This yielded 226 relevant articles. After downloaded and read full articles, only compared results between EPID and CBCT treatment verifications were selected. This yielded, a final total of 23 articles to be included for this work. Conclusion: EPID and MV-CBCT have almost equal potential to indicate the set-up error but since they used the LINAC’s energy, no cross-calibration procedure is needed because the imaging geometry is the same as the treatment geometry though both of them have poor image quality. The image quality of KV-CBCT is superior to both EPID and MV-CBCT and is better to indicate the set-up error. CBCT in conjunction with energy fluence maps from EPID could be used to verify the dose delivered. In this instance, a cross-calibration between the KV CBCT and EPID would be required which would not be necessary with MV CBCT. Though MV CBCT handles inhomogeneity better, soft tissue visualization is superior with KV CBCT. EPID is used to calibrate the LINAC machine. Generally, both EPID and CBCT have their own different advantages and need further study to solve their limitations.

Upgrade to premium to download articles

Sign up to access the full text

Already have an account?login

similar resources

Mega Voltage Cone Beam Computed Tomography (MV- CBCT) using a Standard Medical Linear Accelerator and EPID: A feasibility study.

Introduction: The success of radiotherapy cancer treatment delivery depends on the accuracy of patient positioning for each treatment session. A number of kilovoltage x-ray volumetric imaging modalities with an additional source and detector have been developed to allow patient set-up verification based on the internal anatomy, but a significant portion of medical linacs are on...

full text

Pretreatment quality control of single isocenter half- beam treatment planning technique using an amorphous silicon electronic portal-imaging device (EPID)

Introduction: Electronic portal imaging devices (EPIDs) are fundamentally used for instantaneous verification of the patient set‐up, block shape, and leaf positions during radiation therapy. In radiotherapy, situations arise in which an inclined PTV must be treated mutually with adjacent nodal regions. This methodology is most widely used for matching tangential/lateral breas...

full text

Artifact reduction techniques in Cone Beam Computed Tomography (CBCT) imaging modality

Introduction: Cone beam computed tomography (CBCT) was introduced and became more common based on its low cost, fast image procedure rate and low radiation dose compared to CT. This imaging modality improved diagnostic and treatment-planning procedures by providing three-dimensional information with greatly reduced level of radiation dose compared to 2D dental imaging modalitie...

full text

Cone-Beam Computed Tomography (CT) for a Megavoltage Linear Accelerator (LINAC) Using an Electronic Portal Imaging Device (EPID) and the Algebraic Reconstruction Technique (ART)

This study investigates the feasibility and applicability of cone beam CT for a megavoltage therapeutic LINAC. A Rando head phantom was irradiated using the 15 MV beam of a Varian 2100C LINAC at MSKCC for gantry angles from -102° to 102° with a 2° increment. The projection image for each gantry angle was obtained using a Varian Mark II EPID. Reference images without phantom were also collected ...

full text

Comparison of electronic portal imaging and cone beam computed tomography for position verification in children.

AIM To compare the accuracy of radiotherapy set-up using an electronic portal imaging device (EPID) versus megavoltage cone beam computed tomography (MV-CBCT) in paediatric patients. MATERIALS AND METHODS In total, 204 pairs of EPID and MV-CBCT were carried out for 72 patients in the first 3 treatment days and weekly thereafter. RESULTS For the whole group, the mean systematic EPID set-up e...

full text

Quality Assurance (QA) for Kilovoltage Cone Beam Computed Tomography (CBCT)

Precise targeting of the tumor is very important and imaging can be immensely helpful in this process. Hence the quality assurance of the imaging systems used to guide the beam is also essential. One could argue that it is as critical as quality assurance for the radiation dose output, since the right amount of radiation dose delivered to the wrong location can be as detrimental as delivering t...

full text

My Resources

Save resource for easier access later

Save to my library Already added to my library

{@ msg_add @}


Journal title

volume 15  issue Special Issue-12th. Iranian Congress of Medical Physics

pages  321- 321

publication date 2018-12-01

By following a journal you will be notified via email when a new issue of this journal is published.

Hosted on Doprax cloud platform doprax.com

copyright © 2015-2023