Calculation of the Scattered Radiation Profile in 64 Slice CT Scanners Using Experimental Measurement

Authors

  • Afshin Akbarzadeh Ph.D. Student, Department of Medical Physics and Biomedical Engineering and Research Center for Science and Technology in Medicine, Tehran University of Medical Sciences, Tehran, Iran
  • Hossein Ghadiri Ph.D. Student, Department of Medical Physics and Biomedical Engineering and Research Center for Science and Technology in Medicine, Tehran University of Medical Sciences, Tehran, Iran
  • Mohammad Rreza Ay Assistant Professor in Biomedical Engineering, Biomedical Systems & Medical Physics Dept., Tehran University of Medical Sciences & Research Center for Science and Technology in Medicine, Tehran, Iran.
  • Saied Sarkar Associate Professor in Biomedical Engineering, Biomedical Systems & Medical Physics Dept., Tehran University of Medical Sciences & Research Center for Science and Technology in Medicine, Tehran, Iran.
Abstract:

Introduction: One of the most important parameters in x-ray CT imaging is the noise induced by detected scattered radiation. The detected scattered radiation is completely dependent on the scanner geometry as well as size, shape and material of the scanned object. The magnitude and spatial distribution of the scattered radiation in x-ray CT should be quantified for development of robust scatter correction techniques. Empirical methods based on blocking the primary photons in a small region are not able to extract scatter in all elements of the detector array while the scatter profile is required for a scatter correction procedure. In this study, we measured scatter profiles in 64 slice CT scanners using a new experimental measurement. Material and Methods: To measure the scatter profile, a lead block array was inserted under the collimator and the phantom was exposed at the isocenter. The raw data file, which contained detector array readouts, was transferred to a PC and was read using a dedicated GUI running under MatLab 7.5. The scatter profile was extracted by interpolating the shadowed area. Results: The scatter and SPR profiles were measured. Increasing the tube voltage from 80 to 140 kVp resulted in an 80% fall off in SPR for a water phantom (d=210 mm) and 86% for a polypropylene phantom (d = 350 mm). Increasing the air gap to 20.9 cm caused a 30% decrease in SPR. Conclusion: In this study, we presented a novel approach for measurement of scattered radiation distribution and SPR in a CT scanner with 64-slice capability using a lead block array. The method can also be used on other multi-slice CT scanners. The proposed technique can accurately estimate scatter profiles. It is relatively straightforward, easy to use, and can be used for any related measurement.

Upgrade to premium to download articles

Sign up to access the full text

Already have an account?login

similar resources

calculation of the scattered radiation profile in 64 slice ct scanners using experimental measurement

introduction: one of the most important parameters in x-ray ct imaging is the noise induced by detected scattered radiation. the detected scattered radiation is completely dependent on the scanner geometry as well as size, shape and material of the scanned object. the magnitude and spatial distribution of the scattered radiation in x-ray ct should be quantified for development of robust scatter...

full text

A new method for experimental characterisation of scattered radiation in 64-slice CT scanner

PURPOSE The consummate 64-slice CT scanner that spawns a new generation of non-invasive diagnostic tool, however revolutionary, brings with it the incidental by-product that is scattered radiation. The extended detector aperture capability in the 64-slcie CT scanner allows the effects of scattered radiation to be more pronounced and therefore demands that the magnitude and spatial distribution ...

full text

Multi-Slice CT Scanners

In the second half of 1998, four manufacturers (GE, Siemens, Toshiba and Picker, now Philips) extended this concept by launching multi-slice CT scanners. All were 3rd generation helical scanners, with low voltage slip rings, capable of acquiring four CT slices in one x-ray tube rotation (see Fig. 1). Additional four-slice models have been introduced by Toshiba and GE. Some basic specifications ...

full text

Evaluation of MRI-based Polymer Gel Dosimetry for Measurement of CT Dose Index (CTDI) on 64 slices CT Scanners

Introduction: Computed tomography (CT) has numerous applications in clinical procedures but its main problem is its high radiation dose to the patients compared to other imaging modalities using x-ray. CT delivers approximately high doses to the nearby tissues due to the scattering effect, fan beam (beam divergence) and limited collimator efficiency. The radiation dose from multi-slice scanners...

full text

64-slice scanners build case for coronary CTA

Research focused on the causes, diagnosis, treatment, and prevention of cardiovascular disease is progressing rapidly.1 Radiology has always played a central diagnostic role, particularly in the coronary vessels. A technique for selectively catheterizing the coronary arteries was developed in 1959, and conventional x-ray coronary angiography is still considered the gold standard for detecting c...

full text

Beam profile assessment in spectral CT scanners

In this paper, we present a method that uses a combination of experimental and modeled data to assess properties of x-ray beam measured using a small-animal spectral scanner. The spatial properties of the beam profile are characterized by beam profile shape, the angular offset along the rotational axis, and the photon count difference between experimental and modeled data at the central beam ax...

full text

My Resources

Save resource for easier access later

Save to my library Already added to my library

{@ msg_add @}


Journal title

volume 6  issue 2

pages  1- 10

publication date 2009-06-01

By following a journal you will be notified via email when a new issue of this journal is published.

Hosted on Doprax cloud platform doprax.com

copyright © 2015-2023