design and implementation of a complementary treatment planning software for the gzp6 hdr brachytherapy system (gzp6 ctps)
Authors
abstract
introduction: brachytherapy is one of the most common treatment modalities for gynecological cancer. the gzp6 brachytherapy system is one of the devices utilized in iran. it has been considered particularly due to its low cost compared to other more complete and established systems. this system has some deficiencies including lack of a treatment planning software for non-predefined treatments, inability to change the gradually changeable dosimetric variables and using a point source estimation in dose calculation. this report presents a complementary treatment planning software (ctps) to the system’s own dedicated program. material and methods: first, the dosimetric characteristics of three gzp6 sources were calculated based on the tg-43 protocol using the mcnp4c monte carlo code. then, the calculated dose distribution around the implanted applicators, based on the selected dwell positions and dwell times, was shown in a graphical user interface (gui) written using the matlab software. results: the computation uncertainty in the resulting tg-43 parameters was about 1% and the calculated parameters were in good agreement with similar studies on cobalt-60 source dosimetry. furthermore, the gui is prepared as a user-friendly executable file which can be installed on any operating system. discussion and conclusion: since different patients have distinct anatomy and physical conditions, a program for non-predefined situations of source arrangement is necessary. using gzp6 ctps can satisfy this requirement.
similar resources
Design and Implementation of a Complementary Treatment Planning Software for the GZP6 HDR Brachytherapy System (GZP6 CTPS)
Introduction: Brachytherapy is one of the most common treatment modalities for gynecological cancer. The GZP6 brachytherapy system is one of the devices utilized in Iran. It has been considered particularly due to its low cost compared to other more complete and established systems. This system has some deficiencies including lack of a treatment planning software for non-predefined treatments, ...
full textRadial dose functions of GZP6 intracavitary brachytherapy 60Co sources: treatment planning system versus Monte Carlo calculations
Background: The Monte Carlo (MC) method is not only used for dose calculations around brachytherapy sources but also for benchmarking treatment planning systems (TPS) calculations. Materials and Methods: Three 60Co sources of GZP6 brachytherapy unit were simulated using MCNP4C MC Code. The radial dose functions were calculated by MC method and GZP6 TPS and were compared. Results: There was a go...
full textDose distribution verification for GZP6 sources: a comparison of Monte Carlo, radiochromic film, and GZP6 treatment planning system
www.onk.ns.ac.rs/Archive Vol 20, No. 1–2, July 2012 INTRODUCTION The GZP6 high dose rate (HDR) afterloading brachytherapy unit (Nuclear Power Institute of China) (1) has recently been introduced in the Iranian radiation oncology centers. The unit has six different cobalt-60 source braids being loaded in tandem and ovoid applicators. The sources are used in the treatment of cervix, vagina, rectu...
full textradial dose functions of gzp6 intracavitary brachytherapy 60co sources: treatment planning system versus monte carlo calculations
background: the monte carlo (mc) method is not only used for dose calculations around brachytherapy sources but also for benchmarking treatment planning systems (tps) calculations. materials and methods: three 60co sources of gzp6 brachytherapy unit were simulated using mcnp4c mc code. the radial dose functions were calculated by mc method and gzp6 tps and were compared. results: there was a go...
full textDesign and Implementation of a Software System for Detecting Orthographical or Morphological Errors in Persian Words
This paper presents a new method for analyzing words in the Persian language context to find orthographical and structural errors regardless of the meaning. This technique tokenizes each word in a statement then tries to detect the kind of word, and analyses its correctness in terms of orthography and morphology by means of a lexicon. It should be noted that some words in the Persian language h...
full textAir kerma strength characterization of a GZP6 Cobalt-60 brachytherapy source.
BACKGROUND Task group number 40 (TG-40) of the American Association of Physicists in Medicine (AAPM) has recommended calibration of any brachytherapy source before its clinical use. GZP6 afterloading brachytherapy unit is a (60)Co high dose rate (HDR) system recently being used in some of the Iranian radiotherapy centers. AIM In this study air kerma strength (AKS) of (60)Co source number thre...
full textMy Resources
Save resource for easier access later
Journal title:
iranian journal of medical physicsجلد ۸، شماره ۱، صفحات ۹-۱۸
Hosted on Doprax cloud platform doprax.com
copyright © 2015-2023