Examination of some performance characteristics of breath alcohol measurements obtained with the intoxilyzer 8000C following social drinking conditions.

نویسندگان

  • James H Watterson
  • Kayla N Ellefsen
چکیده

The Intoxilyzer 8000C was used to measure breath alcohol concentration (BrAC) in 10 healthy subjects under social drinking conditions. Measurements commenced within 5 min of the end of drinking (EOD). For 14 blood-breath pairs, measured BrACs were compared to corresponding venous blood alcohol concentrations (vBAC) of samples drawn at least 30 min after EOD and within 5 min of the corresponding breath test. BAC was analyzed using an enzymatic method. Concentration differences between breath and blood (BrAC - vBAC) ranged from -32 to +3 mg/dL (untruncated BrAC) and from -32 to -4 mg/dL (truncated BrAC). The Invalid Sample message was actuated in five out of 23 BrAC profiles. In the remaining 18 samples, residual mouth alcohol was evaluated by comparing the maximum difference between successive (5 min apart) measurements (MID5) over 20-30 min after EOD with the precision of replicate BrAC values taken 30-40 min after EOD (5 mg/dL or less; precision unaffected by breath sample volume over the range of 2-3 L). MID5 values occurred within the first three measurements in 16/18 cases, indicative of a significant mouth alcohol effect. Thus, mandatory delays should be used with the Intoxilyzer 8000C prior to testing to minimize the probability of overestimation of BrAC due to mouth alcohol.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Assessment of response of the Intoxilyzer 8000C to volatiles of forensic relevance in vitro, part I: acetone, isopropanol, and methanol.

The response of the Intoxilyzer 8000C (version approved for evidentiary breath alcohol testing in Canada) to volatile solvents in vitro is described. Acetone, isopropanol, and methanol were prepared as aqueous solutions or dilutions of standard alcohol solution (SAS; 1.21 mg ethanol/mL) to generate apparent blood ethanol concentrations (aBEC) of 50 or 80 mg/dL. Solvent concentrations examined w...

متن کامل

Comparison of ethanol concentrations in venous blood and end-expired breath during a controlled drinking study.

Concentration-time profiles of ethanol were determined for venous whole blood and end-expired breath during a controlled drinking experiment in which healthy men (n=9) and women (n=9) drank 0.40-0.65 g ethanol per kg body weight in 20-30 min. Specimens of blood and breath were obtained for analysis of ethanol starting at 50-60 min post-dosing and then every 30-60 min for 3-6 h. This protocol fu...

متن کامل

A large-scale study of the relationship between blood and breath alcohol concentrations in New Zealand drinking drivers.

Blood alcohol concentrations (BAC) and corresponding breath alcohol concentrations (BrAC) were determined for 21,582 drivers apprehended by New Zealand police. BAC was measured using headspace gas chromatography, and BrAC was determined with Intoxilyzer 5000 or Seres Ethylometre infrared analysers. The delay (DEL) between breath testing and blood sampling ranged from 0.03 to 5.4 h. BAC/BrAC rat...

متن کامل

Lack of Effect of Over-the-Counter Oral Care Strips on a Forensic Breath Alcohol Testing Device

There have been a considerable number of cases litigating the validity of the forensic breath alcohol test and the potentials for a false positive ethanol reading. Recently, a claim asserted that the prior use of an over-the-counter (OTC) “oral care strip” normally used to suppress halitosis of the oral cavity, produced a false positive result when a subject was given a breath alcohol test duri...

متن کامل

Evidenzer – A New Breath-Alcohol Analyzer For Evidential Testing

Background Between 1989 and 2002 the only approved instrument for evidential breath-alcohol testing (EBT) in Sweden was Intoxilyzer 5000S (1). During this time about 160,000 EBTs were performed with 150 Intoxilyzer instruments located throughout the country (1,2). When a new generation of EBTs was being considered the police authorities desired that the instrument should be functional both at a...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of analytical toxicology

دوره 33 8  شماره 

صفحات  -

تاریخ انتشار 2009