Evaluation of MBJ20® Transcutaneous Bilirubinometer in the Assessement of Severity of Neonatal Jaundice
نویسندگان
چکیده
The objective of this study was to evaluate the MBJ20® transcutaneous bilirubinonometer by Beijing M&B Electronic Instrument Co. Ltd. (Beijing, China) in the assessment of severity of neonatal jaundice. Two hundred and twenty-two paired samples from 88 newborns aged between 28 and 44 weeks of gestation at Nnamdi Azikiwe University Teaching Hospital, Nnewi Anambra State, were analyzed. These infants had total serum bilirubin (TSB) levels measured on clinical indication, and transcutaneous bilirubin (TcB) levels were obtained at the forehead and sternum within 10 min of the TSB levels. There was a close correlation between TSB and TcBF and between TSB and TcBS for all the paired readings (n = 222, r = 0.904; n = 222, r = 0.917, respectively). MBJ20 values differed on average from the TSB values by 3.2 mg/dL for TcBF and 2.0 mg/dL for TcBs. We conclude that TcB measurements using the MBJ20 jaundice meter correlate very closely with TSB levels over the range of TSB encountered in this study and that estimation of TcB should be done at both the forehead and sternum.
منابع مشابه
The Value of Bilicheck® as a Screening Tool for Neonatal Jaundice in the South of Iran
The gold standard to assess jaundice in neonates is the serum bilirubin measurement. Blood sampling for the determination of total serum bilirubin (TSB) is painful for newborns and stressful for parents. The Bilicheck®, a new transcutaneous bilirubinometer, is considered as a more accurate measurement of bilirubin compared to the previous bilirubinometers courtesy of its advanced technology. Th...
متن کاملA Comparison between Transcutaneous Bilirubin (TcB) and Total Serum Bilirubin (TSB) Measurements in Term Neonates
Background: Transcutaneous bilirubinometry (TCB) is a simple method for estimating bilirubin levels in neonates. This method is noninvasive, quick, and painless. We aimed to compare serum and cutaneous bilirubin measurements in term neonates.Method: In this descriptive cross-sectional study, 200 neonates with icter with birth weights of at least 2500 grams were studied. TCB was measured using a...
متن کاملTranscutaneous bilirubinometer in assessment of neonatal jaundice in northern India.
This study was undertaken from April 2002 to March 2003 to find out the correlation of transcutaneous bilirubinometer index with serum bilirubin levels in term, pre-term, small for gestation age babies, with and without phototherapy in neonates with jaundice. Another aim was to evaluate the transcutaneous bilirubinometer as a screening device for neonatal hyperbilirubinemia by finding the actio...
متن کاملTrans-Cutaneous Bilirubinometery versus Serum Bilirubin in Neonatal Jaundice.
Hyperbilirubinemia is a common problem in neonates and causes serious complications. Thus, serial measurements of bilirubin should be done. This assessment is done through two methods of laboratory measurement in serum sample and transcutaneous bilirubinometer. This descriptive study compared transcutaneous bilirubin assessment and laboratory serum bilirubin. Bilirubin level was assessed among ...
متن کاملEvaluation of the Severity and Duration of Thrombocytopenia following Exchange Transfusion in neonatal hyperbilirubinemia
Background: Infant jaundice is one the most common causes of hospitalization in infant in the first month of birth, which is defined an abnormal increase in blood bilirubin levels. Exchange transfusion is the recommended treatment for neonatal jaundice who do not respond to phototherapy and experience dangerous complication of jaundice and signs of kernicterus. However, this treatment may lead ...
متن کامل