Recurrent balloon rupture during Transcatheter Aortic Valve Replacement (TAVR) - a challenge to the Hybrid team.
نویسندگان
چکیده
Results We were successful in deploying a 26 mm Edwards Sapien XT through trans-aortic approach after two failed attempt due to rupture of the balloon during valvuloplasty. The patient also had an off pump coronary artery bypass grafting to the right coronary artery. The patient was discharged home after a week. Discussion/Conclusion TAVR balloon rupture is an uncommon complication; it could result when the balloon expands against a sharp spike of calcium as it was in this case. The issue of retrieving the balloon could be complicated in such scenarios particularly if the access was through a transfemoral route. In such a situation the retrieval of the balloon and sheath system would require pulling back the delivery system into the descending aorta and performing a potentially complicated unplanned ilio-femoral vascular exploration. On the other hand, the appearances of sharp spike-like calcifications would favour the trans-aortic approach where retrieval of the system if the balloon ruptures is more straightforward. Pre-operative Computed tomography is very essential in selecting the best access for trans catheter valve.
منابع مشابه
Transcatheter Aortic Valve Replacement: A Review Article
Transcatheter aortic valve replacement (TAVR) is a novel therapeutic intervention for the replacement of severely stenotic aortic valves in high-risk patients for standard surgical procedures. Since the initial PARTNER trial results, use of TAVR has been on the rise each year. New delivery methods and different valves have been developed and modified in order to promote the minimally invasive p...
متن کاملAnatomical and procedural features associated with aortic root rupture during balloon-expandable transcatheter aortic valve replacement.
BACKGROUND Aortic root rupture is a major concern with balloon-expandable transcatheter aortic valve replacement (TAVR). We sought to identify predictors of aortic root rupture during balloon-expandable TAVR by using multidetector computed tomography. METHODS AND RESULTS Thirty-one consecutive patients who experienced left ventricular outflow tract (LVOT)/annular/aortic contained/noncontained...
متن کاملTranscatheter and Doppler waveform correlation in transcatheter aortic valve replacement
Transcatheter aortic valve replacement (TAVR) has become the preferred therapy for treatment of severe aortic stenosis in patients at intermediate to high risk of perioperative mortality following surgical aortic valve replacement. Haemodynamic assessment is an integral part of the procedure, and it is crucial for the operator to have an in-depth understanding of the haemodynamic alterations th...
متن کاملCT image-based engineering analysis of transcatheter aortic valve replacement.
DESPITE THE INCREASED GLOBAL EXPERIENCE WITH TRANSCATHETER AORTIC VALVE REPLACEMENT (TAVR), there remain 3 major adverse events. Aortic rupture (Fig. 1), coronary artery obstruction (Fig. 2), and paravalvular leakage (PVL) (Fig. 3) may occur with valve implantation. Oversizing or excessive radial expansion force with the TAVR stent may cause aortic rupture, whereas insufficient dilation may lea...
متن کاملBioprosthetic Valve Fracture Improves the Hemodynamic Results of Valve-in-Valve Transcatheter Aortic Valve Replacement.
BACKGROUND Valve-in-valve (VIV) transcatheter aortic valve replacement (TAVR) may be less effective in small surgical valves because of patient/prosthesis mismatch. Bioprosthetic valve fracture (BVF) using a high-pressure balloon can be performed to facilitate VIV TAVR. METHODS AND RESULTS We report data from 20 consecutive clinical cases in which BVF was successfully performed before or afte...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of cardiothoracic surgery
دوره 10 Suppl 1 شماره
صفحات -
تاریخ انتشار 2015