Evaluation of the adaptation of the glucose oxidase-peroxidase-3-methyl-2-benzothiazolinone hydrazone-N,N-dimethylaniline procedure to the Technicon "SMA 12-60," and comparison with other automated methods for glucose.

نویسندگان

  • R N Carey
  • D Feldbruegge
  • J O Westgard
چکیده

We have evaluated the Technicon SMA 12/60 modification of a glucose oxidase/peroxidase-3-methyl2-benzothiazolinone hydrazone-N, N-dimethylaniline method, and find it acceptable. Added bilirubin, creatinine, dextrans, and uric acid did not interfere. We compared values for patients’ sera by this method to those by the glucose oxidase/peroxidase-2,2’-azinodiethylbenzothiazoline-6-sulfonic acid, neocuproine, o-toluidine, ferricyanide, and hexokinase methods. For comparisons to the hexokinase method (which we used as a reference method), n = 371, slope = 1.00, y-intercept = 1.57 mg/dl, and bias = 2.05 mg/dl. For the method being evaluated, the run-torun average monthly standard deviation was 2.9 mg/dl for a control product for which the mean was 85 mg of glucose per deciliter, and 4.3 mg/dl for a product for which the mean was 240 mg/dl. All six methods were compared for effects of uremic sera, icteric sera, and sera from patients receiving ascorbic acid therapy or hypoglycemic drugs.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of the BMC glucose oxidase/peroxidase-4-aminophenazone-phenol procedure for glucose as adapted to the Technicon SMAC.

We evaluated the analytical performance of Trinder's glucose oxidase (EC 1.1.3.4)/peroxidase (EC 1.11.1.7) 4-aminophenazone-phenol method for the quantification of serum glucose as adapted to the Technicon SMAC. Our results correlated well with those by the routine SMAC glucose oxidase/peroxidase 3-methyl-2-benzothiazolinone hydrazone-N,N-dimethylaniline method (y = 1.02x - 49.4; r = 0.99) and ...

متن کامل

Evaluation of the co-immobilized hexokinase/glucose-6-phosphate dehydrogenase method for glucose, as adapted to the Technicon SMAC.

We assessed the analytical performance of the co-immobilized hexokinase (EC 2.7.1.1) and glucose-6-phosphate dehydrogenase (EC 1.1.1.49) method for D-glucose analysis on the Technicon SMAC. The enzyme-containing coils were usable for one month, or 12 000 tests. Bilirubin, hemoglobin, lipemia, creatinine, uric acid, citric acid, and ascorbic acid did not interfere. Results with this method were ...

متن کامل

Application of a new peroxide indicator reaction to the specific, automated determination of glucose with glucose oxidase.

We describe an automated, colorimetric determi. nationof glucose in biological fluids that combines the specificityof glucose oxidase and of a new peroxide indicator reaction. In the presence of peroxidase, 3-methyl-2-benzothiazoli none hydrazone oxidativelycouples with N,N-dimethylaniline to form a stable,intensely colored, water-soluble indamine dye, the concentration of which is proportional...

متن کامل

SpectrophotometricDeterminationof UrinaryOxalatewith OxalateOxidase

A novel spectrophotometric enzymlc procedure for estimating oxalic acid in urine is described. Oxalate oxidase, prepared from moss species, converts oxalic acid to hydrogen peroxide and carbon dioxide. Hydrogen peroxide is determined enzymatically with horseradish peroxidase, by oxidative coupling of 3-methyl-2-benzothiazolinone hydrazone with N,N-dimethylaniline; the resulting indamine dye is ...

متن کامل

Evaluation of a glucose oxidase-peroxidase method adapted to the single-channel autoanalyzer and SMA 12-60.

The glucose oxidase-peroxidase method for determining serum glucose, with the ammonium salt of 2,2’-azine-di (3 ethyl benzothiazoline-(6)-sulfonic acid) as the chromogen, was adapted to use with the single-channel AutoAnalyzer and the SMA 12/60 (Technicon Corp.). The extended linearity (up to 500 mg/100 ml on both systems) eliminated the need for numerous dilutions in the 300 to 500 mg/100 ml r...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Clinical chemistry

دوره 20 5  شماره 

صفحات  -

تاریخ انتشار 1974