Evaluation of a dual-level four-bank micro multileaf collimator in reducing dose undulation, leakage, and transmission
نویسندگان
چکیده
This study was to evaluate the reduction of dose undulation at the field edge, gap leakage and interleaf transmission by utilizing a dual-level four-bank micro multileaf collimator (mMLC) and compare with standard MLC. Method and Materials: A 6MV Varian 600C linear accelerator has been collimated with a dual-level fourbank mMLC. The dual mMLC consists of four banks of leaves with upper and lower levels. The two levels of leaves are perpendicular to the central axis and travel orthogonally to each other. It provides the capability of precise aperture shaping and high edge resolution by using micro leaves with widths of 2.1 mm and operating with two levels of leaves. The maximum field size of the dual mMLC is 97 mm x 108 mm formed by 96 leaves. To evaluate the reduction of dose undulation, dose leakage and transmission, radiographic films of a right triangle aperture generated by the two levels dual mMLC were compared with the aperture using only one level of leaves and the other level of leaves remained open. The parameters for comparison include 2D and 3D dose difference, isodose lines, gamma analysis, and dose profiles. Results: The dual mMLC smoothes the dose undulation at the field edge and reduces gap leakage and interleaf transmission. The maximum dose reduction at the field edge is 39%. The gap leakage was reduced from 12% to less than 3.5%. The mean interleaf transmission was reduced from 6% to 3%. Conclusion: This innovative dual mMLC has shown the advantage of reducing dose undulation, gap leakage and interleaf transmission. It offers the ability of better field-edge shaping, reduced gap leakage and interleaf transmission which are considered important characteristics in IMRT and SRT for target conformity and high dose delivery.
منابع مشابه
Physical and Dosimetric Aspect of Euromechanics Add-on Multileaf Collimator on Varian Clinac 2100 C/D
Background:Â Before treatment planning and dose delivery, quality assurance of multi-leaf collimator (MLC) has an important role in intensity-modulated radiation therapy (IMRT) due to the creation of multiple segments from optimization process.Objective: The purpose of this study is to assess the quality control of MLC leaves using EBT3 Gafchromic films.Material and Methods: Leaf Position accur...
متن کاملDosimetric characteristics of dual‐layer multileaf collimation for small‐field and intensity‐modulated radiation therapy applications
The purpose of the present work was to measure the performance characteristics in the penumbra region and on the leaf-end of an innovative dual-layer micro multileaf collimator (DmMLC). The DmMLC consists of two orthogonal (upper and lower) layers of leaves; a standard MLC consists of one layer. The DmMLC provides unique performance characteristics in smoothing dose undulation, reducing leaf-en...
متن کاملBeam characteristic of P.E collimators Add-on multileaf collimator
Introduction: Multileaf collimators (MLCs) are computer-controlled devices which are Useful and cost-effective tools for conformal therapy and intensity modulation radiotherapy (IMRT). Nowadays MLCs are important and standard configuration of the new medical linear accelerators which are substitute of conventional Blocks and has eliminated the ...
متن کاملRadiation transmission, leakage and beam penumbra measurements of a micro‐multileaf collimator using GafChromic EBT film
Micro-multileaf collimator systems coupled to linear accelerators for radioneurosurgery treatments require a rigorous dosimetric characterization in order to be used in 3D conformal and intensity modulated stereotactic radiosurgery and radiotherapy applications. This characterization involves high precision measurements of leaf transmission, leakage and beam penumbra through the collimation sys...
متن کاملMonte Carlo Study of Unflattened Photon Beams Shaped by Multileaf Collimator
Introduction: This study investigates basic dosimetric properties of unflattened 6 MV photon beam shaped by multileaf collimator and compares them with those of flattened beams.Materials and Methods: Monte Carlo simulation model using BEAM code was developed for a 6MV photon beam based on Varian Clinic 600 unique performance linac operated with and without a flattening filter in beam line. Dosi...