Bipolar head design: inner bearing range of motion and disassociation.
نویسندگان
چکیده
To address the clinical problems of joint stiffness, acetabular pain, and component wear, recent bipolar heads have been designed to achieve increased inner bearing range of motion. We tested four designs to determine if this compromises component integrity. Inner bearing ranges of motion were determined and the components then mechanically tested to determine inner bearing pull-out disassociation strengths as well as static and dynamic impingement forces for disassociation. Inner bearing ranges of motion with a 22 mm head were between 65 degrees to 84 degrees for the four prostheses. Pull-out forces for disassociation ranged between 700 N to 1475 N; static impingement forces were 20 Nm to 49 Nm and dynamic impingement forces were 5 Nm to 24 Nm. There was no relation between bipolar head inner bearing range of motion and the potential for component disassociation; however, one design modification produced a lowered disassociation strength. Design modifications must be evaluated by a variety of test methods to adequately determine their effects on bipolar head integrity.
منابع مشابه
Late liner disassociation of a Pinnacle system acetabular component.
This article describes a case of late locking mechanism failure and disassociation of a Pinnacle acetabular cup (DePuy, Warsaw, Indiana) and Marathon polyethylene liner (DePuy) 53 months after routine primary total hip arthroplasty in an active patient. Following an uncomplicated initial postoperative recovery, the patient felt a pop while kneeling for gardening activities. The patient had no p...
متن کاملDoes the self-centering mechanism of bipolar hip endoprosthesis really work in vivo?
PURPOSES To examine radiographically the component motion in a bipolar prosthesis and to determine whether the self-centering mechanism really works in vivo. METHODS 38 patients with 41 bipolar hip endoprostheses (30 for coxarthrosis and 11 for osteonecrosis of femoral head) were included in this study. Two radiographs of each case were taken to evaluate the self-centering mechanism. The firs...
متن کاملاستئوتومی انتهای فوقانی ران برای درمان استئوآرتریت مفصل ران
During the past 20 years, poximal femoral osteotomy has been forgotten because most orthopedic surgeon believed that total hip arthroplasty is the only treatment method for hip joint osteoarthritis. The disadvatages of proximal femoral osteotomy include: Uncertain results, prolonged rehabilitation time after surgery, little improvement in joint ROM and according tosome.euthors difficult subsequ...
متن کاملA new test for bipolar prosthesis disassociation.
Because of several failures by dissociation of a redesigned bipolar prosthesis, a new, dynamic test was developed. This dynamic cam-out test represents a closer simulation of one possible clinical mechanism of bipolar disassociation. The results of dynamic testing are affected by the bipolar design, particularly the locking mechanism that was the problem with the redesigned prosthesis.
متن کاملAssessment of Damage on a Dual Mobility Acetabular System.
BACKGROUND Dual mobility designs were introduced to increase stability and reduce the risk of dislocation, both being common reasons for surgical revision after total hip arthroplasty. The in vivo behavior of dual mobility constructs remains unclear, and to our knowledge, no data have been published describing in vivo surface damage to the polyethylene bearing surfaces. METHODS We used surfac...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Bulletin (Hospital for Joint Diseases (New York, N.Y.))
دوره 63 1-2 شماره
صفحات -
تاریخ انتشار 2005