Directions and magnitudes of misregistration of CT attenuation-corrected myocardial perfusion studies: incidence, impact on image quality, and guidance for reregistration.

نویسندگان

  • John A Kennedy
  • Ora Israel
  • Alex Frenkel
چکیده

UNLABELLED CT-based attenuation-corrected (AC) myocardial perfusion imaging (MPI) studies may show significant artifacts caused by misregistration between SPECT and CT data. The present study aimed at identifying the directions and magnitudes of misregistration with greatest impact on AC myocardial perfusion image quality. METHODS The incidence, magnitude, and direction of misregistration were assessed in 248 consecutive stress-rest MPI studies in 124 patients. In addition, cardiac SPECT/CT registration was artificially modified in 40 studies, shifting CT data by +/-1, +/-2, and +/-3 pixels along the cephalad/caudal, dorsal/ventral, and left/right axes. Percentage of change in 5-wall AC-MPI polar map scoring was calculated for each region, and the direction of the shift along each axis was analyzed statistically (Student t test, P < 0.05) and compared to determine the region most significantly affected by each shift (Newman-Keuls test, P < 0.05). Changes in the normal and abnormal summed stress score (SSS) due to artificial misregistration were assessed (kappa-statistics, McNemar differences). RESULTS SPECT/CT misregistration of more than 1 pixel was found in 73% (181/248) of studies and more than 2 pixels in 23% of studies (57/248). A 3-pixel ventral shift most significantly affected polar map scoring (-15.4% +/- 6.1% change in lateral wall; -7.5% +/- 5.5% change in inferior wall). A 3-pixel dorsal shift resulted in a -9.5% +/- 5.3% apical and -8.8% +/- 5.8% septal change. Polar map scoring was least affected by the cephalad/caudal shift (<5% average change in all regions except for the anterior wall; -9.9% +/- 7.4% change for 3-pixel caudal shift). The most significant changes occurred in the lateral and anterior walls when the myocardium on SPECT overlapped lung tissue on CT, encountered in 16% of studies (40/248). Clinically significant changes (in SSS) occurred for the 3-pixel caudal, dorsal, and right shifts. CONCLUSION A misregistration of significant magnitude occurred in 23% of studies and in the direction of the most severe artifacts in 16% of studies. Severe misregistration along the dorsal/ventral axis most significantly affected AC-MPI. Quality control of SPECT and CT registration and manual realignment should be performed routinely, with the highest priority in AC studies showing an overlay of the myocardium on SPECT with lung tissue on the CT component of AC-MPI.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Attenuation correction in myocardial perfusion SPECT/CT: effects of misregistration and value of reregistration.

UNLABELLED The accuracy of myocardial perfusion SPECT improves with attenuation correction. Algorithms for attenuation correction in hybrid SPECT/CT systems have the potential for misregistration of emission and transmission scans because CT and SPECT are obtained sequentially. Misregistration will influence regional tracer distribution and may reduce diagnostic accuracy. This study focused on ...

متن کامل

The influence of misregistration between CT and SPECT images on the accuracy of CT-based attenuation correction of cardiac SPECT/CT imaging: Phantom and clinical studies

Introduction: Integration of single photon emission computed tomography (SPECT) and computed tomography (CT) scanners into SPECT/CT hybrid systems permit detection of coronary artery disease in myocardial perfusion imaging (MPI). Misregistration between CT and emission data can produce some errors in uptake value of SPECT images. The aim of this study was evaluate the influence...

متن کامل

Evaluation of attenuation correction process in cardiac SPECT images

  Introduction: Attenuation correction is a useful process for improving myocardial perfusion SPECT and is dependent on activity and distribution of attenuation coefficients in the body (attenuation map). Attenuation artifacts are a common problem in myocardial perfusion SPECT. The aim of this study was to compare the effect of attenuation correction using different a...

متن کامل

Artifacts from misaligned CT in cardiac perfusion PET/CT studies: frequency, effects, and potential solutions.

UNLABELLED CT-based attenuation correction is a widely used option in commercial PET/CT scanners. However, as a result of a nonsimultaneous acquisition and differences in temporal resolution between both modalities, a potential misregistration between the PET and CT, especially in the thorax and the upper abdomen, can be found. We observed a substantial number of apparent perfusion defects in s...

متن کامل

Attenuation correction in myocardial perfusion SPECT using sequential transmission - emission scanning with 99mTc [Persain]

Introduction: Nowadays, Imaging of the myocardial perfusion (MPI) using the single photon emission tomography (SPET) in the diagnosis of coronary artery disease, especially myocardial ischemia, is of great importance. In contrast to the coronary artery angiography, MPI is non-invasive, less expensive and more physiological. Unfortunately, this image is affected by the some artifacts. Thes...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of nuclear medicine : official publication, Society of Nuclear Medicine

دوره 50 9  شماره 

صفحات  -

تاریخ انتشار 2009