Reducing Radiation Dose in Coronary Angiography and Angioplasty Using Image Noise Reduction Technology.

نویسندگان

  • Mirlind Kastrati
  • Lukas Langenbrink
  • Michal Piatkowski
  • Jochen Michaelsen
  • Doris Reimann
  • Rainer Hoffmann
چکیده

This study sought to quantitatively evaluate the reduction of radiation dose in coronary angiography and angioplasty with the use of image noise reduction technology in a routine clinical setting. Radiation dose data from consecutive 605 coronary procedures (397 consecutive coronary angiograms and 208 consecutive coronary interventions) performed from October 2014 to April 2015 on a coronary angiography system with noise reduction technology (Allura Clarity IQ) were collected. For comparison, radiation dose data from consecutive 695 coronary procedures (435 coronary angiograms and 260 coronary interventions) performed on a conventional coronary angiography system from October 2013 to April 2014 were evaluated. Patient radiation dosage was evaluated based on the cumulative dose area product. Operators and operator practice did not change between the 2 evaluated periods. Patient characteristics were collected to evaluate similarity of patient groups. Image quality was evaluated on a 5-grade scale in 30 patients of each group. There were no significant differences between the 2 evaluated groups in gender, age, weight, and fluoroscopy time (6.8 ± 6.1 vs 6.9 ± 6.3 minutes, not significant). The dose area product was reduced from 3195 ± 2359 to 983 ± 972 cGycm(2) (65%, p <0.001) in coronary angiograms and from 7123 ± 4551 to 2431 ± 1788 cGycm(2) (69%, p <0.001) in coronary interventions using the new noise reduction technology. Image quality was graded as similar between the evaluated systems (4.0 ± 0.7 vs 4.2 ± 0.6, not significant). In conclusion, a new x-ray technology with image noise reduction algorithm provides a substantial reduction in radiation exposure without the need to prolong the procedure or fluoroscopy time.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of Patients’ Exposure during Angiography and Angioplasty Procedures in the Angiography Department of Shahid Madani Hospital in Tabriz

Introduction: Coronary angiography and angioplasty procedures lead to significant radiation exposure of patients. In the current study, the average radiation dose to patients during angiography in the Angiography Department of Shahid-Madani Hospital was determined. Materials and Methods: An image intensifier based angiography unit (Philips BH 3000) was used for angiography procedures. The accur...

متن کامل

Breast Radiation Protection in Coronary Angiography

Introduction: Coronary angiography (CA) is an interventional diagnostic procedure to assess cardiovascular disease. CA is performed using dedicated fluoroscopic systems equipped with either flat detector or image intensifier. There is an increase in the number of CA annually. Reducing the radiation exposure in the CA is essential, especially due to high fluoroscopy time. In ord...

متن کامل

Assessment of patients skin dose undergoing coronary angiography and Percutaneous Transluminal Coronary Angioplasty (PTCA)

Background: Practice of interventional cardiology procedures such as Coronary angiography (CA) and percutaneous transluminal coronary angioplasty (PTCA) has largely increased in recent years and is now changing to a matter of routine in many hospitals. The purpose of this study was to measure skin and organ doses in patients undergoing CA and PTCA and to find a probable relation with c...

متن کامل

Low kilovolt “prospective ECG-triggering” vs. “retrospective ECG-gating” coronary CTA: comparison of image quality and radiation dose

Background: To compare image quality and radiation doses of low kilovolt (kV) “prospective ECG-triggering” (PT) and standard “retrospective ECG-gating” (RG) coronary computed tomography (CT) angiography. Materials and Methods: A total of 101 consecutive patients (76 males, 25 females; mean age: 55.44 ± 8.28 years) with low-to-intermediate risk status for coronary artery disease and with a body ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The American journal of cardiology

دوره 118 3  شماره 

صفحات  -

تاریخ انتشار 2016