Bench Comparative Assessment of Mechanically Assisted Cough Devices.
نویسندگان
چکیده
BACKGROUND Mechanically assisted cough devices are used in patients with impaired cough to avoid secretion accumulation. We compared 5 mechanically assisted cough devices by bench testing using a breathing simulator and assessed their user-friendliness. METHODS We measured inspiratory and expiratory airway pressures and peak expiratory flow, the strongest indicator of cough efficacy. We performed 2 bench tests: 1) to ascertain the differences between preset and actual settings in 3 different machines of each mechanically assisted cough device and 2) to assess the effects of varying respiratory impedance and air leaks on performance of the devices. We also evaluated the user-friendliness of the devices by measuring the time required and errors in accomplishing 4 tasks by 10 physicians unfamiliar with mechanically assisted cough devices compared with product specialists from the distributing companies. Physicians also scored the ease of use. RESULTS Four mechanically assisted cough devices during insufflation and all 5 during exsufflation showed differences between preset and actual airway pressures. All but one device showed uneven actual pressure values between models of the same type. Peak expiratory flow was significantly influenced by the mechanical properties in 2 devices and by air leaks in 4 devices. The median time to accomplish all tasks by the product specialist (10 [interquartile range of 2-29] s) was overall significantly shorter compared with all physicians (from 19 [14-65] to 36 [19-116] s). The number of procedural errors, but not the perceived ease of use, differed significantly between the devices. CONCLUSIONS The performance of different mechanically assisted cough devices was erratic and included variance between models from the same manufacturer; it was affected by respiratory system impedance and air leaks. Time and rate of errors for performing procedures were elevated. These findings indicate that the devices are not interchangeable and that the settings should be targeted for each patient with the specific machine being used. Improvements in reliability, performance, and user-friendliness are advisable.
منابع مشابه
Duchenne muscular dystrophy: continuous noninvasive ventilatory support prolongs survival.
OBJECTIVE To describe survival outcomes with noninvasive ventilation (NIV) for full ventilatory support, and a mechanically assisted cough and oximetry protocol in a series of patients with Duchenne muscular dystrophy. METHODS We monitored end-tidal carbon dioxide (P(ETCO(2))), S(pO(2)), vital capacity, maximum insufflation capacity, and cough peak flow. Nocturnal NIV was initiated for sympto...
متن کاملExpiratory peak flow and respiratory system resistance in mechanically ventilated patients undergoing two different forms of manually assisted cough.
OBJECTIVE Mechanical ventilation is associated with retained airway secretions. Manually assisted cough contributes to the displacement of bronchial mucus, whereas positive end-expiratory pressure increases collateral ventilation and maintains airway patency. This study aimed to assess the effects of manually assisted cough, either alone or added to increased positive end-expiratory pressure an...
متن کاملEffect of High-Frequency Oscillations on Cough Peak Flows Generated by Mechanical In-Exsufflation in Medically Stable Subjects With Amyotrophic Lateral Sclerosis.
BACKGROUND Mechanically assisted coughing with mechanical in-exsufflation (MI-E) is recommended for noninvasive management of respiratory secretions in amyotrophic lateral sclerosis (ALS). To improve the effectiveness of the technique, a new device combining MI-E with high-frequency oscillations (HFO) has been developed. This work aimed to assess the effect of HFO on the cough peak flow generat...
متن کاملComparative assessment of several automatic CPAP devices' responses: a bench test study
Automatic continuous positive airway pressure (APAP) devices adjust the delivered pressure based on the breathing patterns of the patient and, accordingly, they may be more suitable for patients who have a variety of pressure demands during sleep based on factors such as body posture, sleep stage or variability between nights. Devices from different manufacturers incorporate distinct algorithms...
متن کاملDuchenne muscular dystrophy: survival by cardio-respiratory interventions.
We describe survival in Duchenne dystrophy by invasive and noninvasive ventilation vs. untreated. Patients were untreated prior to 1984 (Group 1), underwent tracheotomy from 1984 until 1991 (Group 2), and were managed by noninvasive mechanical ventilation and cardioprotective medications subsequently (Group 3). Symptoms, vital capacity, and blood gases were monitored for all and spirometry, cou...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Respiratory care
دوره 60 7 شماره
صفحات -
تاریخ انتشار 2015