Reduction of applicator displacement in MR/CT-guided cervical cancer HDR brachytherapy by the use of patient hover transport system

نویسندگان

  • Megan Andrew
  • Yusung Kim
  • Timothy Ginader
  • Brian J. Smith
  • Wenqing Sun
  • Dongxu Wang
چکیده

Purpose To quantify the reduction of relative displacement between the implanted intracavitary applicator and the patient bony anatomy, due to the use of a hover transport system during the patient transports between the imaging table and the treatment table. Material and methods The displacement of the applicator inside the patient was measured by comparing the distance between the tip of the tandem and the pubic bone on X-ray radiography images taken before and after moving a patient to magnetic resonance/computed tomography imaging. Displacements were evaluated for 27 fractions of treatment using hover transport and 185 fractions of treatment using manual transport. Results The use of hover transport system reduced the percentage of fractions with displacements greater than 5 mm from 22.7% to 7.4%. The reduction of applicator displacement using hover transport is statistically significant, compared to the manual transport method (p-value 0.0086; mean displacement 3.41 mm [95% CI: 2.96-3.97] for manual transport, and 2.27 mm [95% CI: 1.71-2.97] for hover transport fractions). Conclusions This study indicates that the hover transport system is effectively reducing displacement between tandem and patient bony anatomy during patient transports. The potential improvement in dosimetric accuracy due to this reduction warrants further study.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Accuracy Evaluation of Oncentraâ„¢ TPS in HDR Brachytherapy of Nasopharynx Cancer Using EGSnrc Monte Carlo Code

Background: HDR brachytherapy is one of the commonest methods of nasopharyngeal cancer treatment. In this method, depending on how advanced one tumor is, 2 to 6 Gy dose as intracavitary brachytherapy is prescribed. Due to high dose rate and tumor location, accuracy evaluation of treatment planning system (TPS) is particularly important. Common methods used in TPS dosimetry are based on computat...

متن کامل

Comparison the treatment time of cobalt and iridium sources in HDR brachytherapy

Introduction: The use of long-half-life Cobalt source instead of iridium source will be very helpful in HDR brachytherapy for treatment of cervical cancer. The aim of this study was compared the treatment time with cobalt and iridium source in brachytherapy for treatment of cervical cancer using Tandem-ovoid applicator. Materials and Methods: In this study, 19...

متن کامل

Applicator Attenuation Effect on Dose Calculations of Esophageal High-Dose Rate Brachytherapy Using EDR2 Film

Introduction Interaluminal brachytherapy is one of the important methods of esophageal cancer treatment. The effect of applicator attenuation is not considered in dose calculation method released by AAPM-TG43. In this study, the effect of High-Dose Rate (HDR) brachytherapy esophageal applicator on dose distribution was surveyed in HDR brachytherapy. Materials and Methods A cylindrical PMMA phan...

متن کامل

Daily CT measurement of needle applicator displacement during multifractionated high-dose-rate interstitial brachytherapy for postoperative recurrent uterine cancer.

We investigated daily needle applicator displacement during multifractionated high-dose-rate interstitial brachytherapy (HDR-ISBT) for postoperative recurrent uterine cancer. Eight patients with postoperative recurrent uterine cancer received HDR-ISBT with or without external beam radiotherapy using our unique ambulatory technique. To analyze displacement, we obtained daily computed tomography ...

متن کامل

Clinical implementation of MR‐guided vaginal cylinder brachytherapy

We present an institutional experience on the clinical implementation of magnetic resonance (MR)-guided vaginal brachytherapy using commercially available solid applicator models. To test the fidelity of solid applicator models to digitize vaginal cylinder applicators, three datasets were evaluated. The first included 15 patients who were simulated with CT alone. Next, a water phantom was used ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 10  شماره 

صفحات  -

تاریخ انتشار 2018