Goldmann and error correcting tonometry prisms compared to intracameral pressure
نویسندگان
چکیده
BACKGROUND Compare Goldmann applanation tonometer (GAT) prism and correcting applanation tonometry surface (CATS) prism to intracameral intraocular pressure (IOP), in vivo and in vitro. METHODS Pressure transducer intracameral IOP was measured on fifty-eight (58) eyes undergoing cataract surgery and the IOP was modulated manometrically to 10, 20, and 40 mmHg. Simultaneously, IOP was measured using a Perkins tonometer with a standard GAT prism and a CATS prism at each of the intracameral pressures. Statistical comparison was made between true intracameral pressures and the two prism measurements. Differences between the two prism measurements were correlated to central corneal thickness (CCT) and corneal resistance factor (CRF). Human cadaver eyes were used to assess measurement repeatability. RESULTS The CATS tonometer prism measured closer to true intracameral IOP than the GAT prism by 1.7+/-2.7 mmHg across all pressures and corneal properties. The difference in CATS and GAT measurements was greater in thin CCT corneas (2.7+/-1.9 mmHg) and low resistance (CRF) corneas (2.8+/-2.1 mmHg). The difference in prisms was negligible at high CCT and CRF values. No difference was seen in measurement repeatability between the two prisms. CONCLUSION A CATS prism in Goldmann tonometer armatures significantly improve the accuracy of IOP measurement compared to true intracameral pressure across a physiologic range of IOP values. The CATS prism is significantly more accurate compared to the GAT prism in thin and less rigid corneas. The in vivo intracameral study validates mathematical models and clinical findings in IOP measurement between the GAT and CATS prisms.
منابع مشابه
Goldmann tonometer error correcting prism: clinical evaluation
PURPOSE Clinically evaluate a modified applanating surface Goldmann tonometer prism designed to substantially negate errors due to patient variability in biomechanics. METHODS A modified Goldmann prism with a correcting applanation tonometry surface (CATS) was mathematically optimized to minimize the intraocular pressure (IOP) measurement error due to patient variability in corneal thickness,...
متن کاملGoldmann applanation tonometry error relative to true intracameral intraocular pressure in vitro and in vivo
BACKGROUND Goldmann applanation tonometry (GAT) error relative to intracameral intraocular pressure (IOP) has not been examined comparatively in both human cadaver eyes and in live human eyes. Futhermore, correlations to biomechanical corneal properties and positional changes have not been examined directly to intracameral IOP and GAT IOP. METHODS Intracameral IOP was measured via pressure tr...
متن کاملDynamic Contour Tonometry, Tono-Pen XL®, and Goldmann Applanation Tonometry in Comparison to Intracameral Intraocular Pressure (IOP) Measurements in Patients with Corneal Pathologies
Unspecified Posted at the Zurich Open Repository and Archive, University of Zurich ZORA URL: http://doi.org/10.5167/uzh-101903 Published Version Originally published at: Neuburger, M; Grosswendt, J; Lautebach, S; Maier, P; Birnbaum, F; Böhringer, D; Funk, J; Jordan, J F; Reinhard, T (2014). Dynamic Contour Tonometry, Tono-Pen XL®, and Goldmann Applanation Tonometry in Comparison to Intracameral...
متن کاملGoldmann tonometry tear film error and partial correction with a shaped applanation surface
Purpose The aim of the study was to quantify the isolated tear film adhesion error in a Goldmann applanation tonometer (GAT) prism and in a correcting applanation tonometry surface (CATS) prism. Methods The separation force of a tonometer prism adhered by a tear film to a simulated cornea was measured to quantify an isolated tear film adhesion force. Acrylic hemispheres (7.8 mm radius) used a...
متن کاملGoldmann Tonometer Prism with an Optimized Error Correcting Applanation Surface
PURPOSE We evaluate solutions for an applanating surface modification to the Goldmann tonometer prism, which substantially negates the errors due to patient variability in biomechanics. METHODS A modified Goldmann or correcting applanation tonometry surface (CATS) prism is presented which was optimized to minimize the intraocular pressure (IOP) error due to corneal thickness, stiffness, curva...
متن کامل