Goldmann tonometer error correcting prism: clinical evaluation
نویسندگان
چکیده
PURPOSE Clinically evaluate a modified applanating surface Goldmann tonometer prism designed to substantially negate errors due to patient variability in biomechanics. METHODS A modified Goldmann prism with a correcting applanation tonometry surface (CATS) was mathematically optimized to minimize the intraocular pressure (IOP) measurement error due to patient variability in corneal thickness, stiffness, curvature, and tear film adhesion force. A comparative clinical study of 109 eyes measured IOP with CATS and Goldmann prisms. The IOP measurement differences between the CATS and Goldmann prisms were correlated to corneal thickness, hysteresis, and curvature. RESULTS The CATS tonometer prism in correcting for Goldmann central corneal thickness (CCT) error demonstrated a reduction to <±2 mmHg in 97% of a standard CCT population. This compares to only 54% with CCT error <±2 mmHg using the Goldmann prism. Equal reductions of ~50% in errors due to corneal rigidity and curvature were also demonstrated. CONCLUSION The results validate the CATS prism's improved accuracy and expected reduced sensitivity to Goldmann errors without IOP bias as predicted by mathematical modeling. The CATS replacement for the Goldmann prism does not change Goldmann measurement technique or interpretation.
منابع مشابه
Goldmann Tonometer Prism with an Optimized Error Correcting Applanation Surface
PURPOSE We evaluate solutions for an applanating surface modification to the Goldmann tonometer prism, which substantially negates the errors due to patient variability in biomechanics. METHODS A modified Goldmann or correcting applanation tonometry surface (CATS) prism is presented which was optimized to minimize the intraocular pressure (IOP) error due to corneal thickness, stiffness, curva...
متن کاملGoldmann and error correcting tonometry prisms compared to intracameral pressure
BACKGROUND Compare Goldmann applanation tonometer (GAT) prism and correcting applanation tonometry surface (CATS) prism to intracameral intraocular pressure (IOP), in vivo and in vitro. METHODS Pressure transducer intracameral IOP was measured on fifty-eight (58) eyes undergoing cataract surgery and the IOP was modulated manometrically to 10, 20, and 40 mmHg. Simultaneously, IOP was measured ...
متن کاملGoldmann tonometry tear film error and partial correction with a shaped applanation surface
Purpose The aim of the study was to quantify the isolated tear film adhesion error in a Goldmann applanation tonometer (GAT) prism and in a correcting applanation tonometry surface (CATS) prism. Methods The separation force of a tonometer prism adhered by a tear film to a simulated cornea was measured to quantify an isolated tear film adhesion force. Acrylic hemispheres (7.8 mm radius) used a...
متن کاملHave You Tested Your Goldmann Tonometer Today?
Intraocular pressure (IOP) is the only known modifiable risk factor for glaucoma. Besides being a cornerstone in glaucoma management, accurate measurement of IOP forms an important component of routine ophthalmic evaluation. Despite the innovation of various newer tonometers, Goldmann applanation tonometer is still considered the Gold standard for IOP measurement. However, there are several rep...
متن کاملClinical evaluation of the pressure phosphene tonometer in patients with glaucoma.
AIM To evaluate the reliability of the pressure phosphene tonometer in comparison with the Goldmann applanation tonometer. METHODS 45 consecutive patients with glaucoma (78 eyes) participated in the study. Eyes with previous eye surgery, a documented peripheral visual field defect, a refractive error of more than 5 dioptres, and patients who were unable to understand the procedure were exclud...