Junctioning longitudinally adjacent PTVs with Helical TomoTherapy
نویسندگان
چکیده
Irradiation of longitudinally adjacent PTVs with Helical TomoTherapy (HT) may be clinically necessary, for example in treating a recurrent PTV adjacent to a previously-treated volume. In this work, the parameters which influence the cumulative dose distribution resulting from treating longitudinally adjacent PTVs are examined, including field width, pitch, and PTV location. In-phantom dose distributions were calculated for various on- and off-axis cylindrical PTVs and were verified by ion chamber and film measurement. Dose distributions were calculated to cover 95% of the PTV by the prescribed dose (DP) using 25 and 50 mm long HT fields with pitches of either 0.3 or 0.45. These dose distributions where then used to calculate the 3D dose distribution in the junction region between two PTVs. The best junction uniformity was obtained for fields of equal width, with larger fields providing better intra-PTV dose homogeneity than smaller fields. Junctioning fields of different widths resulted in a much larger dose inhomogeneity, but this could be improved significantly by dividing the junction end of the PTV treated with the smaller field into multiple (up to 4) sub-PTVs, with the prescribed dose in each sub-PTV decreasing with proximity to the junction region. This provided a PTV matching with dose homogeneity similar to that achieved when junctioning two PTVs, both irradiated by the 50 mm field, and provided a distribution where 95% of the PTV received at least the prescribed dose, with maximum excursions from prescribed dose varying from -19% to +13%. We conclude that junctioning adjacent PTVs is possible. Treating longitudinally adjacent PTVs with different widths is a challenge, but dose uniformity is improved by breaking PTVs into multiple contiguous sub-PTVs modified to feather (broaden) the effective junctioning region.
منابع مشابه
The helical tomotherapy thread effect.
Inherent to helical tomotherapy is a dose variation pattern that manifests as a "ripple" (peak-to-trough relative to the average). This ripple is the result of helical beam junctioning, completely unique to helical tomotherapy. Pitch is defined as in helical CT, the couch travel distance for a complete gantry rotation relative to the axial beam width at the axis of rotation. Without scattering ...
متن کاملComparison of Helical Tomotherapy and Direct Tomotherapy in Bilateral Whole Breast Irradiation in a Case of Bilateral Synchronous Grade 1 and Stage 1 Breast Cancer
BACKGROUND Synchronous bilateral breast cancer is rare. A case is presented where whole breast irradiation (WBI) was planned after breast conserving surgery in a patient with synchronous bilateral breast cancer. A comparison was made between the feasibility of helical tomotherapy and direct tomotherapy. CASE REPORT A 60-year-old woman was found to have bilateral breast nodules on routine mammog...
متن کامل“The optimization of superficial planning target volumes (PTVs) with helical tomotherapy”. [JACMP, 15 (6), 2014]
This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. © 2017 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine. DOI: 10.1002/acm2.12117
متن کاملAbstractID: 8177 Title: A Clinical Example of the Helical Tomotherapy Thread Effect: Robustness Discussion
ID: 8177 Title: A Clinical Example of the Helical Tomotherapy Thread Effect: Robustness Discussion Purpose: A previous study explained why pitches of 0.86/n, where n is an integer, will provide the best junctioning of a helically delivered and diverging beam. The previous treatment was theoretical and based on unmodulated examples. In this study, a clinically relevant case of a palliative lung ...
متن کاملDose–volume histogram comparison between static 5-field IMRT with 18-MV X-rays and helical tomotherapy with 6-MV X-rays
We treated prostate cancer patients with static 5-field intensity-modulated radiation therapy (IMRT) using linac 18-MV X-rays or tomotherapy with 6-MV X-rays. As X-ray energies differ, we hypothesized that 18-MV photon IMRT may be better for large patients and tomotherapy may be more suitable for small patients. Thus, we compared dose-volume parameters for the planning target volume (PTV) and o...
متن کامل