Online rapid sampling microdialysis (rsMD) using enzyme-based electroanalysis for dynamic detection of ischaemia during free flap reconstructive surgery
نویسندگان
چکیده
We describe an enzyme-based electroanalysis system for real-time analysis of a clinical microdialysis sampling stream during surgery. Free flap tissue transfer is used widely in reconstructive surgery after resection of tumours or in other situations such as following major trauma. However, there is a risk of flap failure, due to thrombosis in the flap pedicle, leading to tissue ischaemia. Conventional clinical assessment is particularly difficult in such 'buried' flaps where access to the tissue is limited. Rapid sampling microdialysis (rsMD) is an enzyme-based electrochemical detection method, which is particularly suited to monitoring metabolism. This online flow injection system analyses a dialysate flow stream from an implanted microdialysis probe every 30 s for levels of glucose and lactate. Here, we report its first use in the monitoring of free flap reconstructive surgery, from flap detachment to re-vascularisation and overnight in the intensive care unit. The on-set of ischaemia by both arterial clamping and failure of venous drainage was seen as an increase in lactate and decrease in glucose levels. Glucose levels returned to normal within 10 min of successful arterial anastomosis, whilst lactate took longer to clear. The use of the lactate/glucose ratio provides a clear predictor of ischaemia on-set and subsequent recovery, as it is insensitive to changes in blood flow such as those caused by topical vasodilators, like papaverine. The use of storage tubing to preserve the time course of dialysate, when technical difficulties arise, until offline analysis can occur, is also shown. The potential use of rsMD in free flap surgery and tissue monitoring is highly promising.
منابع مشابه
Continuous Online Microdialysis Using Microfluidic Sensors: Dynamic Neurometabolic Changes during Spreading Depolarization
Microfluidic glucose biosensors and potassium ion selective electrodes were used in an in vivo study to measure the neurochemical effects of spreading depolarizations (SD), which have been shown to be detrimental to the injured human brain. A microdialysis probe implanted in the cortex of rats was connected to a microfluidic PDMS chip containing the sensors. The dialysate was also analyzed usin...
متن کاملThe role of antioxidants in ischaemia-reperfusion in a human DIEP flap model.
BACKGROUND A better understanding of the pathophysiology of ischaemia-reperfusion injury and a possible treatment for it, is of great importance. The deep inferior epigastric perforator (DIEP) flap is an innovative, clinical model of ischaemia-reperfusion. There has been an ongoing interest in the health benefits and medical applications of antioxidants. We hypothesised that during ischaemia-re...
متن کاملReverse Peroneal Artery Flap for Large Heel and Sole Defects: A Reliable Coverage
BACKGROUND Large soft tissue defects of ankle and foot always have been challenging to reconstruct. Reverse sural flaps, free flaps have been used for this problem with variable success. Reverse peroneal artery flap is an option to use with reliability without microvascular repair. Connections of peroneal artery around talus and ankle joint are deep and reliable with anterior tibial and post...
متن کاملPalate Mucoperosteum: An Usefull Adjunct in Buccal Mucosa Reconstruction
Background: Palate is a complex structure separating oro- and nasopharynx. However, reconstruction of the defects of palate is much simpler because of the versatile mucoperiosteal flaps. Here, we present our experience of palatal mucoperiosteal flap used in different situations. Methods: Fifteen patients of palatal as well as buccal mucosa defects were reconstructed using either free or pedicl...
متن کاملDetection of myocardial ischaemia using surface microdialysis on the beating heart.
Microdialysis (MD) can be used to study metabolism of the beating heart. We investigated whether microdialysis results obtained from epicardial (surface) sampling reflect acute changes in the same way as myocardial sampling from within the substance of the ventricular wall. In anaesthetized open-thorax pigs a coronary snare was placed. One microdialysis probe was placed with the sampling membra...
متن کامل