Large-diameter total hip arthroplasty modular heads require greater assembly forces for initial stability
نویسندگان
چکیده
OBJECTIVES Modular junctions are ubiquitous in contemporary hip arthroplasty. The head-trunnion junction is implicated in the failure of large diameter metal-on-metal (MoM) hips which are the currently the topic of one the largest legal actions in the history of orthopaedics (estimated costs are stated to exceed $4 billion). Several factors are known to influence the strength of these press-fit modular connections. However, the influence of different head sizes has not previously been investigated. The aim of the study was to establish whether the choice of head size influences the initial strength of the trunnion-head connection. MATERIALS AND METHODS Ti-6Al-4V trunnions (n = 60) and two different sizes of cobalt-chromium (Co-Cr) heads (28 mm and 36 mm; 30 of each size) were used in the study. Three different levels of assembly force were considered: 4 kN; 5 kN; and 6 kN (n = 10 each). The strength of the press-fit connection was subsequently evaluated by measuring the pull-off force required to break the connection. The statistical differences in pull-off force were examined using a Kruskal-Wallis test and two-sample Mann-Whitney U test. Finite element and analytical models were developed to understand the reasons for the experimentally observed differences. RESULTS 36 mm diameter heads had significantly lower pull-off forces than 28 mm heads when impacted at 4 kN and 5 kN (p < 0.001; p < 0.001), but not at 6 kN (p = 0.21). Mean pull-off forces at 4 kN and 5 kN impaction forces were approximately 20% larger for 28 mm heads compared with 36 mm heads. Finite element and analytical models demonstrate that the differences in pull-off strength can be explained by differences in structural rigidity and the resulting interface pressures. CONCLUSION This is the first study to show that 36 mm Co-Cr heads have up to 20% lower pull-off connection strength compared with 28 mm heads for equivalent assembly forces. This effect is likely to play a role in the high failure rates of large diameter MoM hips.Cite this article: A. R. MacLeod, N. P. T. Sullivan, M. R. Whitehouse, H. S. Gill. Large-diameter total hip arthroplasty modular heads require greater assembly forces for initial stability. Bone Joint Res 2016;5:338-346. DOI: 10.1302/2046-3758.58.BJR-2016-0044.R1.
منابع مشابه
Does Choice of Head Size and Neck Geometry Affect Stem Migration in Modular Large-Diameter Metal-on-Metal Total Hip Arthroplasty? A Preliminary Analysis
Due to their theoretical advantages, hip systems combining modular necks and large diameter femoral heads have gradually gained popularity. However, among others, concerns regarding changes in the load transfer patterns were raised. Recent stress analyses have indeed shown that the use of modular necks and big femoral heads causes significant changes in the strain distribution along the femur. ...
متن کاملLarge Diameter Femoral Heads Impose Significant Alterations on the Strains Developed on Femoral Component and Bone: A Finite Element Analysis
Total Hip Arthroplasty aims at fully recreating a functional hip joint. Over the past years modular implant systems have become common practice and are widely used, due to the surgical options they provide. In addition Big Femoral Heads have also been implemented in the process, providing more flexibility for the surgeon. The current study aims at investigating the effects that femoral heads of...
متن کاملTaper Failure After Large-Diameter Metal-on-Metal Total Hip Arthroplasty.
The use of metal-on-metal bearing surfaces for total hip arthroplasty increased dramatically in the past decade for a variety of reasons, including the advantage of using larger-diameter femoral heads to reduce instability. However, recent research suggesting negative outcomes and high failure rates associated with some of these implants has decreased their use. Further, the use of larger-diame...
متن کاملPrimary metal-on-metal total hip arthroplasty with large-diameter femoral heads: a clinical trial of 59 hips.
Large-diameter femoral heads with nearly anatomical sizes became available for metal-on-metal total hip arthroplasty after recent advances in metal-on-metal technology. We retrospectively studied the clinical and radiological results in 59 hips of 54 patients (32 women and 22 men, mean age 54.4 years) who underwent cementless metal-on-metal total hip arthroplasty with large-diameter heads. Pati...
متن کاملTotal Hip Arthroplasty with Modular Stem for Dysplastic Hips in South Asian Population
Introduction: Optimum component positioning and orientation is required to optimize the functional result during total hip arthroplasty for dysplastic hips. Patients and methods: Sixty-two patients (66 hips) including 33 males and 29 females underwent total hip arthroplasty using modular stem prosthesis at an average age of 40.6 years (range 17 to 49 years). Nineteen hips were clas...
متن کامل