Clinical Evaluation of the New High-Throughput Luminex NxTAG Respiratory Pathogen Panel Assay for Multiplex Respiratory Pathogen Detection.
نویسندگان
چکیده
A broad range of viral and bacterial pathogens can cause acute respiratory tract infection. For rapid detection of a broad respiratory pathogen spectrum, multiplex real-time PCR is ideal. This study evaluated the performance of the new Luminex NxTAG Respiratory Pathogen Panel (NxTAG-RPP) in comparison with the BioFire FilmArray Respiratory Panel (FA-RP) or singleplex real-time PCR as reference. A total of 284 clinical respiratory specimens and 3 influenza A/H7N9 viral culture samples were tested. All clinical specimens were processed and analyzed in parallel using NxTAG-RPP and the reference standard method. The H7N9 viral culture samples were tested using NxTAG-RPP only. Overall, the NxTAG-RPP demonstrated ≥93% sensitivity and specificity for all respiratory targets except human coronavirus OC43 (HCoV-OC43) and HCoV-HKU1. The H7N9 virus was detected by the influenza A virus matrix gene target, while other influenza A virus subtyping gene targets in the panel remained negative. Complete concordance between NxTAG-RPP and FA-RP was observed in 98.8% (318/322) of positive results (kappa = 0.92). Substantial agreement was found for most respiratory targets, but significant differences were observed in human metapneumovirus (P = 0.001) and parainfluenza virus type 3 (P = 0.031). NxTAG-RPP has a higher sample throughput than FA-RP (96 samples versus 1 sample per run) while the turnaround times for NxTAG-RPP and FA-RP were 5 h (up to 96 samples) and 1 h (for one sample), respectively. Overall, NxTAG-RPP demonstrated good diagnostic performance for most respiratory pathogens. The high sample throughput with reasonable turnaround time of this new assay makes it a suitable multiplex platform for routine screening of respiratory specimens in hospital-based laboratories.
منابع مشابه
Evaluation of NxTAG Respiratory Pathogen Panel and Comparison with xTAG Respiratory Viral Panel Fast v2 and Film Array Respiratory Panel for Detecting Respiratory Pathogens in Nasopharyngeal Aspirates and Swine/Avian-Origin Influenza A Subtypes in Culture Isolates
This study evaluated a new multiplex kit, Luminex NxTAG Respiratory Pathogen Panel, for respiratory pathogens and compared it with xTAG RVP Fast v2 and FilmArray Respiratory Panel using nasopharyngeal aspirate specimens and culture isolates of different swine/avian-origin influenza A subtypes (H2N2, H5N1, H7N9, H5N6, and H9N2). NxTAG RPP gave sensitivity of 95.2%, specificity of 99.6%, PPV of 9...
متن کاملComparison of Luminex NxTAG Respiratory Pathogen Panel and xTAG Respiratory Viral Panel FAST Version 2 for the Detection of Respiratory Viruses
Owing to advancements in molecular diagnostics, recent years have seen an increasing number of laboratories adopting respiratory viral panels to detect respiratory pathogens. In December 2015, the NxTAG respiratory pathogen panel (NxTAG RPP) was approved by the United States Food and Drug Administration. We compared the clinical performance of this new assay with that of the xTAG respiratory vi...
متن کاملEvaluation of the Luminex xTAG Respiratory Viral Panel FAST v2 assay for detection of multiple respiratory viral pathogens in nasal and throat swabs in Vietnam
BACKGROUND Acute respiratory infections (ARI) are among the leading causes of hospitalization in children ≤5 years old. Rapid diagnostics of viral pathogens is essential to avoid unnecessary antibiotic treatment, thereby slowing down antibiotic-resistance. We evaluated the diagnostic performance of the Luminex xTAG Respiratory Viral Panel FAST v2 against viral specific PCR as reference assays f...
متن کاملSite-specific clinical evaluation of the Luminex xTAG gastrointestinal pathogen panel for detection of infectious gastroenteritis in fecal specimens.
We evaluate the clinical performance of the Luminex xTAG gastrointestinal (GI) pathogen in vitro diagnostic (IVD) assay in a comparison between clinical and public health laboratories. The site reproducibility study showed 98.7% sensitivity with high positive and negative agreement values (96.2% and 99.8%, respectively), while assay performance against confirmatory methods resulted in 96.4% sen...
متن کاملDevelopment of Multiplex Reverse Transcription-Polymerase Chain Reaction for Simultaneous Detection of Influenza A, B and Adenoviruses
Background and objective:Millions of people in developing countries lose their lives due to acute respiratory infections, such as Influenza A & B and Adeno viruses. Given the importance of rapid identification of the virus, in this study the researchers attempted to design a method that enables detection of influenza A, B, and adenoviruses, quickly and simultaneously. ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of clinical microbiology
دوره 54 7 شماره
صفحات -
تاریخ انتشار 2016