Composition and morphology of wear debris in failed uncemented total hip replacement.
نویسندگان
چکیده
Interfacial membranes collected at revision from 11 failed uncemented Ti-alloy total hip replacements were examined. Particles in the membranes were characterised by electron microscopy, microchemical spectroscopy and particle size analysis. Most were polyethylene and had a mean size of 0.53 micron +/- 0.3. They were similar to the particles seen in the base resin used in the manufacture of the acetabular implants. Relatively few titanium particles were seen. Fragments of bone, stainless steel and silicate were found in small amounts. Most of the polyethylene particles were too small to be seen by light microscopy. Electron microscopy and spectroscopic techniques are required to provide an accurate description of this debris.
منابع مشابه
Wear particles and ions from cemented and uncemented titanium-based hip prostheses—A histological and chemical analysis of retrieval material
Wear debris-induced inflammation is considered to be the main cause for periprosthetic osteolysis in total hip replacements (THR). The objective of this retrieval study was to examine the tissue reactions and exposure to metal ions and wear particles in periprosthetic tissues and blood samples from patients with titanium (Ti)-based hip prostheses that were revised due to wear, osteolysis, and/o...
متن کاملResults of uncemented porous-coated anatomic total hip replacement.
We reviewed a consecutive series of 241 uncemented, porous-coated anatomic (PCA) hip replacements at an average follow-up of five years (2 to 9). Of these, 32 had failed (13%), 26 at the acetabular component (11%) and six at the femoral component (2%). Acetabular failure was associated with local osteolysis and excessive polyethylene wear in 20 cases: in these histological examination showed gi...
متن کاملA conservative femoral replacement for total hip arthroplasty. A prospective study.
Between 1985 and 1993, 146 patients (162 hips) had total hip replacement (THR) using a conservative uncemented femoral component. The mean age of the patients was 50.8 years and the mean follow-up was 6.2 years (2 to 13). One patient was lost to follow-up, one died within two years of surgery and one had a revision procedure after a fracture sustained in a road-traffic accident. For the remaini...
متن کاملFemoral stem wear in cemented total hip replacement.
The great success of cemented total hip replacement to treat patients with endstage osteoarthritis and osteonecrosis has been well documented. However, its long-term survivorship has been compromised by progressive development of aseptic loosening, and few hip prostheses could survive beyond 25 years. Aseptic loosening is mainly attributed to bone resorption which is activated by an in-vivo mac...
متن کاملFocal femoral osteolysis after revision hip replacement with a cannulated, hydroxyapatite-coated long-stemmed femoral component: a new route for particulate wear debris.
We describe a case of symptomatic focal femoral osteolysis around a screw hole distal to the hydroxyapatite-coated portion of a cannulated femoral component in a revision hip replacement. No locking screw had been inserted into this, the most proximal of the three distal holes for locking screws. The presence of polyethylene wear debris in the tissue excised from the lesion suggested that it ha...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- The Journal of bone and joint surgery. British volume
دوره 76 1 شماره
صفحات -
تاریخ انتشار 1994