Design of a Low-Noise Preamplifier for Nerve Cuff Electrode Recording

نویسندگان

  • Robert Rieger
  • Andreas Demosthenous
  • Peter J. Langlois
چکیده

This paper discusses certain important issues involved in the design of a nerve signal preamplifier for implantable neuroprostheses. Since the electroneurogram signal measured from cuff electrodes is typically on the order of 1 V, a very low-noise interface is essential. We present the argument for the use of BiCMOS technology in this application and then describe the design and evaluation of a complete preamplifier fabricated in a 0.8m double-metal double-poly process. The preamplifier has a nominal voltage gain of 100, a bandwidth of 15 kHz, and a measured equivalent input-referred noise voltage spectral density of 3.3 nV Hz at 1 kHz. The total input-referred rms noise voltage in a bandwidth 1 Hz–10 kHz is 290 nV, the power consumption is 1.3 mW from 2.5-V power supplies, and the active area is 0.3 mm.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Considerations on Improving the Design of Cuff Electrode for ENG Recording - Geometrical Approach, Dedicated IC, Sensitivity and Noise Rejection

Cuff electrodes have several advantages for in situ recording ENG signal. They are easy to implant and not very invasive for the patient. Nevertheless, they are subject to background parasitic noise, especially the EMG generated by the muscles. We show that the use of cuff electrodes with large numbers of poles can to increase their sensitivity and their selectivity with respect to a efficient ...

متن کامل

Considerations on Improving the Design of CUFF Electrode for ENG Recording - Geometrical Approach, Dedicated IC, Sensitivity, Noise Rejection

Cuff electrodes have several advantages for in situ recording ENG signal. They are easy to implant and not very invasive for the patient. Nevertheless, they are subject to background parasitic noise, especially the EMG generated by the muscles. We show that the use of cuff electrodes with large numbers of poles can to increase their sensitivity and their selectivity with respect to a efficient ...

متن کامل

Design of a Peripheral Nerve Electrode for Improved Neural Recording of the Cervical Vagus Nerve by Bita Sadeghlo

Design of a Peripheral Nerve Electrode for Improved Neural Recording of the Cervical Vagus Nerve Bita Sadeghlo Master of Applied Sciences Graduate Department of Electrical & Computer Engineering University of Toronto 2013 Vagus nerve stimulation (VNS) is an approved therapy for patients suffering from refractory epilepsy. While VNS is currently an open loop system, making the system closed loop...

متن کامل

Signal-to-noise Ratio of Nerve Signals Recorded with Full and Open Cylinder Cuff Electrodes

Ab-& In this paper, we compared the signal-to-noise ratios (SNRs) of nerve recordings obtained with an open cuff (a 270" cylinder) and a full cylindrical electrode (simulated by lifting the open electrode outside the solution) using an in vitro set-up. The thermal noise was considered as the main source of noise. The SNR with the open cuff was only 52*7% (n=3) less than that of a full cylindric...

متن کامل

Whole Nerve Recordings with the Spiral Nerve Cuff Electrode

The feasibility of whole nerve recordings from the hypoglossal (HG) nerve is demonstrated in acute cats using the spiral nerve cuff electrode. A good contact between the nerve and the electrodes, provided by the spiral nerve cuff due to its self-coiling property, should improve the signal-tonoise ratio. An instrumentation amplifier with very low input noise characteristics is also utilized. The...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2001