15-nW Biopotential LPFs in 0.35-µm CMOS Using Subthreshold-Source-Follower Biquads With and Without Gain Compensation
نویسندگان
چکیده
Most biopotential readout front-ends rely on the g m- C lowpass filter (LPF) for forefront signal conditioning. A small g m realizes a large time constant ( τ = C / g m) suitable for ultra-low-cutoff filtering, saving both power and area. Yet, the noise and linearity can be compromised, given that each g m cell can involve one or several noisy and nonlinear V- I conversions originated from the active devices. This paper proposes the subthreshold-source-follower (SSF) Biquad as a prospective alternative. It features: 1) a very small number of active devices reducing the noise and nonlinearity footsteps; 2) No explicit feedback in differential implementation, and 3) extension of filter order by cascading. This paper presents an in-depth treatment of SSF Biquad in the nW-power regime, analyzing its power and area tradeoffs with gain, linearity and noise. A gain-compensation (GC) scheme addressing the gain-loss problem of NMOS-based SSF Biquad due to the body effect is also proposed. Two 100-Hz 4th-order Butterworth LPFs using the SSF Biquads with and without GC were fabricated in 0.35- μm CMOS. Measurement results show that the non-GC (GC) LPF can achieve a DC gain of -3.7 dB (0 dB), an input-referred noise of 36 μV rms (29 μV rms ), a HD3@60 Hz of -55.2 dB ( - 60.7 dB) and a die size of 0.11 mm² (0.08 mm²). Both LPFs draw 15 nW at 3 V. The achieved figure-of-merits (FoMs) are favorably comparable with the state-of-the-art.
منابع مشابه
CMOS Cascode Source-Drain Follower for Monolithically Integrated Biosensor Array
Source-drain follower has been designed and implemented for monolithically integrated biosensor array. The circuit acts as a voltage follower, in which a sensing transistor is operated at fixed gatesource and gate-drain voltages. It operates at 10 nW power dissipation. The wide-swing cascode configurations are investigated in constant and non-constant biasing methods. The constant biased cascod...
متن کاملFully Integrated Biopotential Acquisition Analog Front-End IC
A biopotential acquisition analog front-end (AFE) integrated circuit (IC) is presented. The biopotential AFE includes a capacitively coupled chopper instrumentation amplifier (CCIA) to achieve low input referred noise (IRN) and to block unwanted DC potential signals. A DC servo loop (DSL) is designed to minimize the offset voltage in the chopper amplifier and low frequency respiration artifacts...
متن کاملNew Method for Analysis of image sensor to produce and evaluate the image
In this paper, a new method for evaluating CMOS image sensors based on computer modeling and analysis is introduced. Image sensors are composed of different parts, each of which has a specific effect on image quality. Circuits of image sensors can be evaluated and analyzed using circuit simulators or theoretically, but these methods cannot help to produce the final image. In order to produce th...
متن کاملHigh Performance Op Amp Design – Review
In a never-ending effort to reduce power consumption and gate oxide thickness, the integrated circuit industry is constantly developing smaller power supplies. Furthermore, in an effort to reduce costs and integrate analog and digital circuits onto a single chip, MOS devices are used in the weak inversion region or the subthreshold inversion region to minimize dc source power. A feasible config...
متن کاملA Low Power Low Noise Two Stage CMOS Operational Amplifier for Biopotential Signal Acquisition System
Monitoring of the biopotential signals in present scenario attract much attention to the researcher due to increasing demand for better quality of life and to enhance interaction between humans and machines. Biopotential signals are relatively weak amplitude and low frequency in nature. Therefore it is required to design a front circuitry that defines quality of signal for processing to the nex...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- IEEE transactions on biomedical circuits and systems
دوره 7 5 شماره
صفحات -
تاریخ انتشار 2013