Peak Purity Determination with a Diode Array Detector
نویسندگان
چکیده
Peak purity analysis is an evaluation technique for detecting the presence of coeluting impurities in highperformance liquid chromatography (HPLC) data. Running a peak purity check prior to analytical quantitation helps to ensure accuracy. In the development of analytical methods, peak purity analysis can reveal the presence of contamination during standardization and, by doing so, can prevent the subsequent generation of false analytical data. Peak purity analysis is also a useful addition to routine quality control procedures, especially in the analysis of pharmaceuticals and food products, for which contamination and quality of results are critical.
منابع مشابه
Enhanced Diode Array Detector Sensitivity and Automated Peak Purity Control
The introduction of the diode array detector helped overcome this problem by providing access to quantitative as well as qualitative spectral information for a wide wavelength range. Agilent Technologie introduced the first diode array detector for liquid chromatography in 1982. Since that time significant improvements in the design of both the hardware and software have been made, Enhanced Dio...
متن کاملPeak purity analysis in HPLC and CE using diode-array technology Application
In terms of quality control and for all quantitative analysis peak purity is a major task, which can be addressed in different ways. Often only peak shapes and chromatograms are taken into consideration but a very elegant possibility without the need to use a mass spectrometric detector is the comparison of spectra recorded with a diode-array detector during the registration of a chromatographi...
متن کاملValidated stability-indicating HPLC method for the determination of mesna in presence of its degradation products.
A rapid and simple stability-indicating liquid chromatographic method was developed and validated for analysis of mesna in the presence of its degradation products in drug substance and drug products in a run time not >5 min. The separation was achieved on a RP amide C16 column at room temperature using methanol-phosphate buffer (10:90, v/v, pH 3.0) as mobile phase at a flow rate of 1 mL min(-1...
متن کاملSimultaneous RP-HPLC determination of sparfloxacin and dexamethasone in pharmaceutical formulations
The present study describes the development and subsequent validation of simple and accurate stability indicating RP-HPLC method for the determination of sparfloxacin and dexamethasone in pharmaceutical formulations in the presence of their stress-induced degradation products. Both the drugs and their stress-induced degradation products were separated within 10 minutes using C8 column and mixtu...
متن کاملMonte Carlo Determination of Full Energy Peak Efficiency for HPGe Detector and Self-Absorption Correction of Environmental Samples
The Monte Carlo method was used to determine full energy efficiency of a high-purity germanium (HPGe) coaxial detector within an energy range of 53.2-2614 keV. Also, measurement was carried out for a standard Marinelli beaker of 600cm3, which was placed into the reference material of mixed gamma. The plot of the experimentally derived effic...