Transcutaneous estimation of arterial carbon dioxide in intensive care. Which electrode temperature?

نویسندگان

  • G Cheriyan
  • P Helms
  • F Paky
  • D Marsden
  • M C Chiu
چکیده

Transcutaneous and arterial carbon dioxide were measured simultaneously in 57 children (age range 10 days to 14.3 years) undergoing intensive care. All were haemodynamically stable at the time of study. Mean calibration time with 5 and 10% carbon dioxide was 43 (range 38-58) minutes and mean arterialisation time was 10.5 (range 3-30) minutes. Duplicate hourly arterial samples over a four hour period showed that transcutaneous: arterial carbon dioxide correlation was independent of electrode temperatures over the range 42-44 degrees C and was independent of child age. One and a half hours after electrode placement transcutaneous carbon dioxide (kPa) = arterial carbon dioxide X 1.41 + 0.02. Use of a simple ratio (arterial carbon dioxide = transcutaneous carbon dioxide/1.40) resulted in a mean estimate of arterial carbon dioxide within 3% of the actual value with coefficients of variation of 11 and 15% at 1.5 and 4.5 hours, respectively. For prediction of arterial carbon dioxide 95% confidence limits around the regression mean rose from +/- 1.04 kPa (7.8 mm Hg) at 1.5 hours to +/- 1.56 (11.7 mm Hg) at 4.5 hours. Baseline drift was more than +/- 0.67 kPa (5 mm Hg) during 22% of the studies. For a drift of less than 0.67 kPa 95% confidence limits were 0.87 and 1.12 kPa, respectively. Arterial carbon dioxide can be estimated with clinically acceptable tolerances from surface electrodes operating at temperatures between 42-44 degrees C. Improved electrode stability and speed of calibration should make this a valuable monitoring technique.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

End-tidal carbon dioxide and transcutaneous carbon dioxide monitoring during neonatal transport

Abbreviations: CO2, Carbon dioxide; PaCO2, Arterial partial pressure of carbon dioxide; PtcCO2, Transcutaneous partial pressure of carbon dioxide; PetCO2, End-tidal partial pressure of carbon dioxide; NETS, Newborn Emergency Transport Service (Victoria); NICU, Neonatal Intensive Care Unit; PaO2/PAO2 ratio, alveolar-arterial ratio. ADC-FNN Online First, published on April 29, 2005 as 10.1136/adc...

متن کامل

Use of a Combined SpO2/PtcCO2 Sensor in the Delivery Room

Arterial oxygen saturation (SaO(2)) and partial arterial pressure of carbon dioxide (PaCO(2)) are important respiratory parameters in critically ill neonates. A sensor combining a pulse oximeter with the Stow-Severinghaus electrode, required for the measurement of peripheral oxygen saturation (SpO(2)) and transcutaneous partial pressure of carbon dioxide (PtcCO(2)), respectively, has been recen...

متن کامل

Effects of the transcutaneous electrode temperature on the accuracy of transcutaneous carbon dioxide tension

AIM The harmful effect of hypocapnia on the neonatal brain emphasizes the importance of monitoring arterial carbon dioxide tension (PaCO2). Transcutaneous monitoring of carbon dioxide (tcPCO2) reduces the need for arterial blood sampling. Drawbacks are high electrode temperature causing risks of skin burning. The aim was to determine the accuracy and precision of tcPCO2 at reduced electrode tem...

متن کامل

Use of a combined oxygen and carbon dioxide transcutaneous electrode in the estimation of gas exchange during exercise.

BACKGROUND Accurate and reliable measurement of gas exchange during exercise has traditionally involved arterial cannulation. Non-invasive devices to estimate arterial oxygen (O2) and carbon dioxide (CO2) tensions are now available. A method has been devised and evaluated for measuring gas exchange during exercise with a combined transcutaneous O2 and CO2 electrode. METHODS Symptom limited ex...

متن کامل

The accuracy of non-invasive carbon dioxide monitoring: a clinical evaluation of two transcutaneous systems.

We determined the accuracy of two transcutaneous carbon dioxide monitoring systems (SenTec Digital Monitor with V-Sign Sensor and TOSCA 500 with TOSCA Sensor 92) for the measurement of single values and trends in the arterial partial pressure of carbon dioxide in 122 adult patients during major surgery and in 50 adult patients in the intensive care unit. One or several paired measurements were ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Archives of disease in childhood

دوره 61 7  شماره 

صفحات  -

تاریخ انتشار 1986