Biomechanical Comparison of Inter-fragmentary Compression Pressures: Lag Screw versus Herbert Screw for Anterior Odontoid Screw Fixation
نویسندگان
چکیده
OBJECTIVE The purpose of the present study was to compare inter-fragmentary compression pressures after fixation of a simulated type II odontoid fracture with the headless compression Herbert screw and a half threaded cannulated lag screw. METHODS We compared inter-fragmentary compression pressures between 40- and 45-mm long 4.5-mm Herbert screws (n=8 and n=9, respectively) and 40- and 45-mm long 4.0-mm cannulated lag screws (n=7 and n=10, respectively) after insertion into rigid polyurethane foam test blocks (Sawbones, Vashon, WA, USA). A washer load cell was placed between the two segments of test blocks to measure the compression force. Because the total length of each foam block was 42 mm, the 40-mm screws were embedded in the cancellous foam, while the 45-mm screws penetrated the denser cortical foam at the bottom. This enabled us to compare inter-fragmentary compression pressures as they are affected by the penetration of the apical dens tip by the screws. RESULTS The mean compression pressures of the 40- and 45-mm long cannulated lag screws were 50.48±1.20 N and 53.88±1.02 N, respectively, which was not statistically significant (p=0.0551). The mean compression pressures of the 40-mm long Herbert screw was 52.82±2.17 N, and was not statistically significant compared with the 40-mm long cannulated lag screw (p=0.3679). However, 45-mm Herbert screw had significantly higher mean compression pressure (60.68±2.03 N) than both the 45-mm cannulated lag screw and the 40-mm Herbert screw (p=0.0049 and p=0.0246, respectively). CONCLUSION Our results showed that inter-fragmentary compression pressures of the Herbert screw were significantly increased when the screw tip penetrated the opposite dens cortical foam. This can support the generally recommended surgical technique that, in order to facilitate maximal reduction of the fracture gap using anterior odontoid screws, it is essential to penetrate the apical dens tip with the screw.
منابع مشابه
An in vitro biomechanical evaluation of an expansive double-threaded bi-directional compression screw for fixation of type II odontoid process fractures
Odontoid process fracture accounts for 5% to 15% of all cervical spine injuries, and the rate is higher among elderly people. The anterior cannulated screw fixation has been widely used in odontoid process fracture, but the fixation strength may still be limited under some circumstances. This study aims to investigate the biomechanical fixation strength of expansive double-threaded bi-direction...
متن کاملBiomechanical Comparison Between Bashti Bone Plug Technique and Biodegradable Screw for Fixation of Grafts in Ligament surgery
Background: Ligament reconstruction is a common procedure in orthopedic surgery. Although several popular techniques are currently in use, new methods are proposed for secure fixation of the tendon graft into the bone tunnel. Purposes: We sought to introduce our new technique of Bashti bone plug for fixation of soft tissue graft in anterior cruciate ligament (ACL) reconstruction and to compar...
متن کاملA new guide tube for odontoid screw fixation for unstable odontoid fractures: report of 6 case series.
AIM We describe a modified form of traditional open surgery with a new guide tube. This guide tube permits anterior screwing of odontoid in a shorter time with a more simple technique as compared to traditional open surgery, endoscopic and percutaneous surgeries. MATERIAL AND METHODS Our series includes 6 patients who were operated for unstable odontoid fracture. We used a new guide tube for ...
متن کاملBiomechanical evaluation of four different posterior screw and rod fixation techniques for the treatment of the odontoid fractures.
Problems that screw cannot be inserted may occur in screw-rod fixation techniques such as Harms technique. We compared the biomechanical stability imparted to the C-2 vertebrae by four designed posterior screw and rod fixation techniques for the management of odontoid fractures. A three-dimensional finite element model of the odontoid fracture was established by subtracting several unit structu...
متن کاملDelayed Pharyngeal Extrusion of an Anterior Odontoid Screw
A 27-year-old woman with a type II odontoid fracture was treated by anterior odontoid screw fixation. Radiographic union at the fracture site was obtained 3 months after surgery. Nearly 3 years after surgery, she presented at a local Ear, Nose, and Throat (ENT) clinic with a 2-month history of dysphagia. Laryngoscopy identified the head of the odontoid lag screw. Plain radiography showed that t...
متن کامل