The impact of electrode resistance on the biogalvanic characterisation technique

نویسندگان

  • J H Chandler
  • D A Head
  • M E Hubbard
  • A Neville
  • D G Jayne
  • P R Culmer
چکیده

Measurement of a tissue-specific electrical resistance may offer a discriminatory metric for evaluation of tissue health during cancer surgery. With a move toward minimally-invasive procedures, applicable contact sensing modalities must be scalable, fast and robust. A passive resistance characterisation method utilising a biogalvanic cell as an intrinsic power source has been proposed as a potentially suitable solution. Previous work has evaluated this system with results showing effective discrimination of tissue type and damage (through electroporation). However, aspects of the biogalvanic cell have been found to influence the characterisation performance, and are not currently accounted for within the system model. In particular, the electrode and salt-bridge resistance are not independently determined, leading to over-predictions of tissue resistivity. This paper describes a more comprehensive model and characterisation scheme, with electrode parameters and salt-bridge resistivity being evaluated independently. In a generalised form, the presented model illustrates how the relative resistive contributions from the electrodes and medium relate to the existing characterisation method efficacy. We also describe experiments with physiologically relevant salt solutions (1.71, 17.1, 154 mM), used for validation and comparison. The presented model shows improved performance over the current biogalvanic measurement technique at the median conductivity. Both the proposed and extant system models become unable to predict conductivity accurately at high conductivity due to the dominance of the electrodes. The characterisation techniques have also been applied to data collected on freshly excised human colon tissue (healthy and cancerous). The findings suggest that the resistance of the cell under the test conditions is electrode dominated, leading to erroneous tissue resistance determination. Measurement optimisation strategies and the surgical applicability of the biogalvanic technique are discussed in light of these findings.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Assessment of electrochemical properties of a biogalvanic system for tissue characterisation

Biogalvanic characterisation is a promising method for obtaining health-specific tissue information. However, there is a dearth of understanding in the literature regarding the underlying galvanic cell, electrode reactions and their controlling factors which limits the application of the technique. This work presents a parametric electrochemical investigation into a zinc–copper galvanic system ...

متن کامل

Characterisation and Corrosion Performance of MultilayerNano Nickel Coatings on AZ31 Magnesium Alloy

Ni-P and Ni layers multilayer coatings were applied to AZ31 magnesium alloy utilizing electroless and electrodeposition procedures. The aim of the project was to decrease cracks and increase corrosion resistance of the coatings. In order to compare the coatings, the effect of single layer electroless Ni-P coatings with different thicknesses was also investigated. The microstructure and phase co...

متن کامل

Sensitive Electrochemical Determination of Gallic Acid: Application in Estimation of Total Polyphenols in Plant Samples

A modified electrode was prepared by modification of the carbon paste electrode (CPE) with graphene nano-sheets. The fabricated modified electrode exhibited an electrocatalytic activity toward gallic acid (GA) oxidation because of good conductivity, low electron transfer resistance and catalytic effect. The graphene modified CPE had a lower overvoltage and enhanced electrical current respect to...

متن کامل

Analysis of Resistance Spot Welding Process Parameters Effect on the Weld Quality of Three-steel Sheets Used in Automotive Industry: Experimental and Finite Element Simulation

In the present research, the effects of spot-welding process parameters on the nugget diameter and electrode penetration depth of spot-welded joints were investigated. To achieve this, a spot-welded joint of three-thin sheet low carbon steels (same thicknesses of 0.8 mm) was simulated as an electerical-thermal-mechanical coupling of 3D finite element model. After validating the finite element s...

متن کامل

Using electrochemical impedance spectroscopy of salicylate anion selective electrode: simulation for behavior of electrode

In this paper, the behavior of the salicylate anion-selective electrode was studied using the electrochemical impedance spectroscopy technique. Considering the diagram of the charge transfer resistance and -logarithm of the double layer capacitance versus -logarithm of the concentration, linear range concentration increased to 1.0×10^{-8}-1.0×10^{-1} M and 1.0×10^{-9}-1.0×10^{-1} M, respectivel...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 38  شماره 

صفحات  -

تاریخ انتشار 2017