Assessment of blood sample stability for complete blood count using the Sysmex XN-9000 and Mindray BC-6800 analyzers
نویسندگان
چکیده
BACKGROUND Different hematological analyzers have different analytical performances that are often reflected in the criteria for sample stability of the complete blood count. This study aimed to assess the stability of several hematological parameters using the XN-9000 Sysmex and BC-6800 Mindray analyzers. METHODS The impact of storage at room temperature and 4°C was evaluated after 2, 4, 6, 8, 24, 36 and 48h using ten normal and 40 abnormal blood samples. The variation from the baseline measurement was evaluated by the Steel-Dwass-Critchlow-Fligner test and by Bland-Altman plots, using quality specifications and critical difference as the total allowable variation. RESULTS Red blood cells and reticulocyte parameters (i.e. hematocrit, mean corpuscular volume, mean corpuscular hemoglobin concentration, red blood cell distribution width, immature reticulocyte fractions, low-fluorescence reticulocytes, middle-fluorescence reticulocytes, high fluorescence mononuclear cells) showed less stability compared to leukocyte and platelet parameters (except for monocyte count and mean platelet volume). The bias for hematocrit, mean corpuscular volume, mean corpuscular hemoglobin concentration and red blood cell distribution width coefficient of variation was higher than the critical difference after 8h using both analyzers. CONCLUSION Blood samples measured with both analyzers do not show analytically significant changes in up to 2h of storage at room temperature and 4°C. However, the maximum time for analysis can be extended for up to 8h when the bias is compared to the critical difference.
منابع مشابه
Immature Granulocyte Count on the New Sysmex Xn9000: Performance and Diagnosis of Sepsis in the Intensive Care Unit
Introduction. The amount of immature leukocytes reflects marrow response to bacterial infection, and this may be quantified as the band or immature granulocyte (IG) count. The aim of this study was to analyze the IG count performance of the Sysmex XN-9000 hematology analyzer in intensive care unit (ICU) patients. Methods. 480 peripheral blood samples from adult patients admitted to the ICU (301...
متن کاملThe New Sysmex XN-2000 Automated Blood Cell Analyzer More Accurately Measures the Absolute Number and the Proportion of Hematopoietic Stem and Progenitor Cells Than XE-2100 When Compared to Flow Cytometric Enumeration of CD34+ Cells
The gold standard to count circulating hematopoietic stem and progenitor cells (HPCs) is to enumerate CD34 cells by using flow cytometry [1, 2]. XE-2100 (Sysmex, Kobe, Japan), a currently used hematology analyzer, can be used to quantify HPCs, generating counts that reportedly correlate with those obtained using flow cytometric analysis of CD34 cells; however, XE-2100 HPC counts were 1.86-fold ...
متن کاملThe Sysmex XN-2000 Hematology Autoanalyzer Provides a Highly Accurate Platelet Count than the Former Sysmex XE-2100 System Based on Comparison with the CD41/CD61 Immunoplatelet Reference Method of Flow Cytometry
Dear Editor The recently launched hematology autoanalyzer Sysmex XN2000 (Sysmex, Kobe, Japan) provides a new method (PLT-F method) for platelet counting based on a fluorescent RNA staining dye named oxazine [1, 2]. The main advantage of the XN2000’s PLT-F method over the former instrument XE-2100’s optical method (PLT-O method) is the ability to identify apoptotic white cells and not count them...
متن کاملComparison of automated differential blood cell counts from Abbott Sapphire, Siemens Advia 120, Beckman Coulter DxH 800, and Sysmex XE-2100 in normal and pathologic samples.
Reliable automated blood cell characterization and quantification remain challenging in pathologic samples, whereas slide reviews due to unnecessary flagging should be avoided. We compared 4 modern hematology analyzers-Abbott Sapphire, Siemens Advia 120, Sysmex XE-2100, and Beckman Coulter DxH 800-regarding complete blood cell count (CBC), leukocyte differential count, and flagging efficacy in ...
متن کاملDifferential blast counts obtained by automated blood cell analyzers.
BACKGROUND Automated blood cell analyzers often read leukemic blasts as normal cells. In this study, we evaluated the 5-part differential patterns of blasts using automated analyzers to determine if they can differentiate among blast types. METHODS Blood samples containing 10% or more blasts were collected from patients with acute leukemia (N=175). The 5-part differential count was conducted ...