Biomechanical evaluation of shape-memory alloy staples for internal fixation—an in vitro study
نویسندگان
چکیده
BACKGROUND The field of orthopaedics is a constantly evolving discipline. Despite the historical success of plates, pins and screws in fracture reduction and stabilisation, there is a continuing search for more efficient and improved methods of fracture fixation. The aim of this study was to evaluate shape-memory staples and to compare them to a currently used implant for internal fracture fixation. Multi-plane bending stability and interfragmentary compression were assessed across a simulated osteotomy using single and double-staple fixation and compared to a bridging plate. METHODS Transverse osteotomies were made in polyurethane blocks (20 × 20 × 120 mm) and repairs were performed with one (n = 6), or two (n = 6) 20 mm nitinol staples, or an eight-hole 2.7 mm quarter-tubular plate (n = 6). A pressure film was placed between fragments to determine contact area and compressive forces before and after loading. Loading consisted of multi-planar four-point bending with an actuator displacement of 3 mm. Gapping between segments was recorded to determine loads corresponding to a 2 mm gap and residual post-load gap. RESULTS Staple fixations showed statistically significant higher mean compressive loads and contact areas across the osteotomy compared to plate fixations. Double-staple constructs were superior to single-staple constructs for both parameters (p < 0.001). Double-staple constructs were significantly stiffer and endured significantly larger loads before 2 mm gap formation compared to other constructs in the dorsoventral plane (p < 0.001). However, both staple constructs were significantly less stiff and tolerated considerably lower loads before 2 mm gap formation when compared to plate constructs in the ventrodorsal and right-to-left lateral loading planes. Loading of staple constructs showed significantly reduced permanent gap formation in all planes except ventrodorsally when compared to plate constructs. CONCLUSIONS Although staple fixations were not as stable as plate fixations in particular loading planes, double-staple constructs demonstrated the most consistent bending stiffness in all planes. Placing two perpendicular staples is suggested instead of single-staples whenever possible, with at least one staple applied on the compression side of the anticipated loading to improve construct stability.
منابع مشابه
Effect of Electrical Current on Nitinol Medical Staples Shape Memory
Medical staples are one of the nitinol shape memory alloys applications which their pedicles have to be bent for suitable function with an appropriate angle at a specified temperature .It is achievable by shape memory effect. For this purpose, samples of nitinol super elastic alloys were bent in a steel mold with different angles & formed orthopedic staples. Then shape memory effect was induced...
متن کاملEffect of Electrical Current on Nitinol Medical Staples Shape Memory
Medical staples are one of the nitinol shape memory alloys applications which their pedicles have to be bent for suitable function with an appropriate angle at a specified temperature .It is achievable by shape memory effect. For this purpose, samples of nitinol super elastic alloys were bent in a steel mold with different angles & formed orthopedic staples. Then shape memory effect was induced...
متن کاملA Biomechanical Research of Growth Control of Spine by Shape Memory Alloy Staples
Shape memory alloy (SMA) staples in nickel titanium with shape memory effect are effective for spinal growth control. This study was designed to evaluate the biomechanical properties of the staples and observe the stability of the fixed segments spine after the staples were implanted. According to the vertical distance of the vertebrae, SMA staples of 5, 6.5, and 8 mm were designed. The recover...
متن کاملAnalysis on Biomechanical Characteristics of Shape Memory Alloy Fixation Grip for Femoral Fractures
As life the expectancy has increased with the advancement of medical technology, the frequency of femur fractures has also increased. The conventional treatment utilizes a bone fixation plate and screw system. However, because the screws are inserted into directly into the bones, it can cause bones to get crushed and leave holes during removal of the screws. For the reasons above, secondary fra...
متن کاملNitinol compression staples for bone fixation in foot surgery.
We reviewed the use of compression staples made of the memory metal alloy nitinol for fixation in foot surgery. A retrospective study was performed of 31 feet in 27 patients who underwent arthrodesis or an osteotomy fixated using compression staples. OSStaples (BME, San Antonio, Texas) were used in 18 feet, and Memodyn staples (Telos Medical, Fallston, Maryland) were used in 13 feet. A total of...
متن کامل