Mechanical and dose delivery accuracy evaluation in radiosurgery using polymer gels
نویسندگان
چکیده
The polymer gel-Magnetic Resonance Imaging (MRI) dosimetry technique was employed to evaluate the mechanical and dose delivery accuracy in Leksell gamma knife stereotactic radiosurgery for the treatment of multiple targets. Two different polymer gel dosimeter formulations that have been reported in the literature were prepared in-house. A plan for the treatment of four brain metastases (targets) was generated. The plan involved the delivery of four, 8 mm collimator shots using different prescription isodose lines and different prescription doses for each target, keeping the maximum dose constant for all the targets. A sample of each gel formulation was irradiated using a custom made phantom with an experimental procedure capable of testing the increased nominal mechanical accuracy of stereotactic radiosurgery. The irradiated dosimeters were evaluated using a clinical 1.5 T Philips ACS NT MR imager. Result manipulation in 3-D allowed for the determination of the mechanical accuracy in the delivery of each shot through the comparison of measured versus planned shot center coordinates. Dose delivery accuracy was also evaluated by comparison of maximum dose values measured at the center of each shot as well as dose distribution measurements, with corresponding treatment planning calculations. Polymer gel dosimetry was found capable of verifying the complete chain of radiosurgery treatment in gamma knife applications involving the irradiation of multiple targets.
منابع مشابه
Verifying the accuracy of dose distribution in Gamma Knife unit in presence of inhomogeneities using PAGAT polymer gel dosimeter and MC simulation
Background: Polymer gel dosimetry is still the only dosimetry method for direct measuring of threedimensional dose distributions. MRI Polymer gel dosimeters are tissue equivalent and can act as a phantom material. In this study the obtained isodose maps with PAGAT polymer gel dosimeter were compared to those calculated with EGSnrs for singleshot irradiations of 8 and 18 mm collimators of...
متن کاملHow to perform dosimetry with Optical CT
Both polymer gels and PRESAGE radiochromic solid dosimeter, in conjunction with optical CT scanning system, have been employed to measure 3-D dose distribution. The 3-D dose maps obtained from these systems can provide a useful tool for dose verification on complex treatments such as IMRT, radiosurgery, and RapidArc. These complex treatments present high dose gradient regions in the boundaries ...
متن کاملDosimetric Evaluation of Linac Photon Small Fields using MAGIC Polymer Gels
Introduction: In radiotherapy, methods of treatment planning are becoming increasingly more complicated. This requires verification of the doses delivered to increasingly smaller and more precise regions. Radiotherapy techniques are continuously employing smaller and smaller field sizes to deliver tighter radiation doses with higher therapeutic ratios, generating interest among researchers to p...
متن کاملOptical Characterization of NIPAM and PAGAT Polymer Gels for Radiation Dosimetry
Introduction The purpose of the current study was to determine optical sensitivity of N-isopropyl acrylamide NIPAM and polyacrylamide gelatin and tetrakis hydroxymethyl phosphoniun chloride (PAGAT) polymer gels for different wavelength of visible light spectrum applied in optical computed tomography method. Materials and Methods NIPAM and PAGAT polymer gels with conventional formulations used f...
متن کاملEvaluation of the Effects of Inhomogeneities on Dose Profiles Using Polymer Gel Dosimeter and Monte Carlo Simulation in Gamma Knife
Introduction Polymer gel dosimeters offer a practical solution to 3D dose verification for conventional radiotherapy as well as intensity-modulated and stereotactic radiotherapy. In this study, EGSnrc calculated and PAGAT polymer gel dosimeter measured dose profiles from single shot irradiation with 18 mm collimator of Gamma Knife in homogeneous and inhomogeneous phantoms were compared with eac...