A dosimetric study of Leipzig applicators.

نویسندگان

  • José Pérez-Calatayud
  • Domingo Granero
  • Facundo Ballester
  • Vicente Puchades
  • Emilio Casal
  • Angela Soriano
  • Vicente Crispín
چکیده

PURPOSE To obtain the absolute dose-rate distribution in liquid water for all six cup-shaped Leipzig applicators by means of an experimentally validated Monte Carlo (MC) code. These six applicators were used in high-dose-rate (HDR) afterloaders with the "classic" and v2 (192)Ir sources. The applicators have an inner diameter of 1, 2, and 3 cm, with the source traveling parallel or perpendicular to the contact surface. METHODS AND MATERIALS The MC GEANT4 code was used to obtain the dose-rate distribution in liquid water for the six applicators and the two HDR source models. To normalize the applicator output factors, a MC simulation for the "classic" and v2 sources in air was performed to estimate the air-kerma strength. To validate this specific application and to guarantee that realistic source-applicator geometry was considered, an experimental verification procedure was implemented in this study, in accordance with the TG43U1 recommendations. Thermolumniscent dosimeter chips and a parallel plate ionization chamber in a polymethyl methacrylate (PMMA) phantom were used to verify the MC results for the six applicators in a microSelectronHDR afterloader with the "classic" source. Dose-rate distributions dependence on phantom size has been evaluated using two different phantom sizes. RESULTS Percentage depth dose and off-axis profiles were obtained normalized at a depth of 3 mm along the central axis for both phantom sizes. A table of output factors, normalized to 1 U of source kerma strength at this depth, is presented. The dose measured in the PMMA phantom agrees within experimental uncertainties with the dose obtained by the MC GEANT4 code calculations. The phantom size influence on dose-rate distributions becomes significant at depths greater than 5 cm. CONCLUSIONS MC-detailed simulation was performed for the Nucletron Leipzig HDR applicators. The matrix data obtained, with a grid separation of 0.5 mm, can be used to build a dataset in a convenient format to model these distributions for routine use with a brachytherapy treatment planning system.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Physical characteristics of electron beam from conventional and beam shaper IOERT applicator: A comparison study

Introduction: Intraoperative electron radiation therapy (IOERT) is one of the cancer treatment techniques that delivers high doses to tumor bed during surgery. IOERT can be performed by either conventional LINACs or dedicated IORT accelerators such as LIAC (Light Intraoperative Accelerator). Two types of applicators can be used with LIAC dedicated accelerator including conventi...

متن کامل

Comparison between small radiation therapy electron beams collimated by Cerrobend and tubular applicators

The purpose of this study was to compare the dosimetric properties of small field electron beams shaped by circular Cerrobend blocks and stainless steel tubular applicators. Percentage depth dose curves, beam profiles, and output factors of small-size circular fields from 2 to 5 cm diameter, obtained either by tubular applicators and Cerrobend blocks, were measured for 6, 10, and 15 MeV electro...

متن کامل

Evaluation of depth dose characteristics of superficial X-rays machine using different kVp and applicators diameter

Background: Radiotherapy Treatment Planning requires different dosimetric quantities as input in order to calculate a desired dose distribution. This study has been focused to evaluate the depth dose characteristics of superficial X-rays being used for radiotherapy treatment. Materials and Methods: Computerized 3-D water phantom of multi-data system was used. The measurements were made...

متن کامل

Dosimetric characteristics of a new unit for electronic skin brachytherapy

PURPOSE Brachytherapy with radioactive high dose rate (HDR) (192)Ir source is applied to small skin cancer lesions, using surface applicators, i.e. Leipzig or Valencia type. New developments in the field of radiotherapy for skin cancer include electronic brachytherapy. This technique involves the placement of an HDR X-ray source close to the skin, therefore combining the benefits of brachythera...

متن کامل

Monte Carlo Simulation of Electron Beams produced by LIAC Intraoperative Radiation Therapy Accelerator

Background: One of the main problems of dedicated IORT accelerators is to determine dosimetric characteristics of the electron beams. Monte Carlo simulation of IORT accelerator head and produced beam will be useful to improve the accuracy of beam dosimetry.Materials and Methods: Liac accelerator head was modeled using the BEAMnrcMonte Carlo simulation system. Phase-space files were generated at...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • International journal of radiation oncology, biology, physics

دوره 62 2  شماره 

صفحات  -

تاریخ انتشار 2005