Phantom validation of 4D flow: independent validation of flow velocity quantification using particle imaging velocimetry
نویسندگان
چکیده
Background 4D phase contrast magnetic resonance blood flow measurements (4D PC-MR) is a promising tool for clinical assessment of cardiac function, but lack of independent validation may limit clinical adoption. Previous validation studies are limited by lack of pulsatility [1,2] or by the need of a post hoc correction factor to compensate for differences in experimental setup between 4D PC-MR and reference imaging [3]. Therefore, we aimed to validate 4D PC-MR velocity measurements with particle
منابع مشابه
In-Vitro Validation of Phase Contrast MRI in a Stenotic Phantom Under Steady Flow Using PIV
Introduction: Blood flow dynamics has an important role in atherosclerosis initiation, progression, and thrombosis leading to occlusive arterial diseases [1]. Since the development of Phase Contrast MRI (PC-MRI), it has been widely used to measure the velocity and flow in blood vessels. Accuracy of the PC-MRI velocity measurements in a 1.5T scanner has previously been evaluated using Laser Dopp...
متن کاملAccuracy of flow measurement with phase contrast MRI in a stenotic phantom: where should flow be measured?
Methods Herein, a series of detailed experiments are reported for validation of MR measurements of steady and pulsatile flows with stereoscopic particle image velocimetry (SPIV) on three different stenotic models with 50%, 74%, and 87% area occlusions. Mean inlet Reynolds number was 190 for both steady and pulsatile cases, mimicking the flow of the human common iliac artery. Axial PC-MRI images...
متن کاملAccuracy and limitations of vector flow mapping: left ventricular phantom validation using stereo particle image velocimetory
BACKGROUND The accuracy of vector flow mapping (VFM) was investigated in comparison to stereo particle image velocimetry (stereo-PIV) measurements using a left ventricular phantom. VFM is an echocardiographic approach to visualizing two-dimensional flow dynamics by estimating the azimuthal component of flow from the mass-conservation equation. VFM provides means of visualizing cardiac flow, but...
متن کاملPhantom validation of 4D flow: independent validation of vortex ring volume quantification using planar laser-induced fluorescence
Cardiac MR group Lund, Dept. of Clinical Physiology, Lund University, Lund, Sweden Full list of author information is available at the end of the article Figure 1 Design and construction of the vortex ring flow phantom. a) 2D view of the flow rectifier and vortex ring tank. All measures are in millimeters (mm). b) 3D view of the flow rectifier and vortex ring tank. c) Pump design. A servo motor...
متن کاملVortex-ring mixing as a measure of diastolic function of the human heart: Phantom validation and initial observations in healthy volunteers and patients with heart failure.
PURPOSE To present and validate a new method for 4D flow quantification of vortex-ring mixing during early, rapid filling of the left ventricle (LV) as a potential index of diastolic dysfunction and heart failure. MATERIALS AND METHODS 4D flow mixing measurements were validated using planar laser-induced fluorescence (PLIF) in a phantom setup. Controls (n = 23) and heart failure patients (n =...