Pacing and sensing interference by air bubble surrounding the non-extended helix of intact active fixation lead.
نویسندگان
چکیده
BACKGROUND Active fixation pacing leads with silicon cylinder tips have been used for their safety and flexibility. Measurement of baseline sensing/pacing characteristics before fixation of helix helps to identify the optimal pacing site, but we encountered difficulties in making these measurements despite multiple attempts with the model LPA 1200M lead. To identify the cause and overcome this complication, we compared 4 different retractable active fixation lead models, which enabled baseline sensing/pacing measurements before extension of helix. METHODS AND RESULTS: We immersed 4 different lead tips and rings in a 0.18% saline solution, and measured the lead impedance before and after flushing of air bubble visible inside the lead tip. Before evacuation of the air bubble, the impedance of the model LPA 1200M lead was >4,000 Ω in 8 out of 10 measurements, although that of the other leads was within the measurable range. After evacuation of the air bubble, the lead impedance returned to within the measurable range. There was no prominent change in the impedance of the metal cylinder tip lead. CONCLUSIONS Air bubbles may interfere with the measurement of baseline sensing/pacing characteristics before active fixation of pacing leads with cylindrical silicon tips. In the case of high impedance beyond the measurable range before extension of helix, the measurement should be repeated after fixation into the myocardium before suspecting lead dysfunction.
منابع مشابه
Morphology Of Current Of Injury Does Not Predict Long Term Active Fixation ICD Lead Performance
BACKGROUND Currents of injury (COI) have been associated with improved lead performance during perioperative measurements in pacemaker and ICD implants. Their relevance on long term lead stability remains unclear. METHODS Unipolar signals were recorded immediately after active fixation ICD lead positioning, blinded to the implanting surgeon. Signals were assigned to prespecified COI types by ...
متن کاملEffectiveness of epicardial atrial pacing using a bipolar steroid-eluting endocardial lead with active fixation in an experimental model
PURPOSE To assess the effectiveness of bipolar epicardial atrial pacing using an active fixation bipolar endocardial lead implanted on the atrial surface in an experimental model. METHODS A total of ten Large White adult pigs underwent pacemaker implantation under general anesthesia. Atrial pacing and sensing parameters were obtained at the procedure, immediate postoperative period and on the...
متن کاملComparison of mid-term clinical experience with steroid-eluting active and passive fixation ventricular electrodes in children.
Although active fixation ventricular leads seem to have advantages over passive fixation leads, this study compares the follow-up results of active and passive fixation leads in children. We evaluated the implantation and follow-up data of 41 children with active (Accufix II DEC, group 1) (n = 20) or passive (Membrane E, group 2) (n = 21) fixation, steroid-eluting ventricular leads. All but one...
متن کاملRight Ventricular Pacing and Sensing Function in High Posterior Septal and Apical Lead Placement in Cardiac Resynchronization Therapy
BACKGROUND The conventional right ventricular (RV) lead position in cardiac resynchronization therapy pacemakers (CRT-P) is the RV apex (RV-A). Little is known about electrophysiological stability and associated complications of pacing leads in RV high posterior septal (RV-HS) position in CRT-P. METHODS Two hundred and thirty-five consecutive CRT-P patients were included from 1999-2010. Pacin...
متن کاملPerformance of Air Foam Flooding under Low Frequency Vibration
Foam injection is widely applied in amounts of fields to drilling, production, and formation protection. Sometimes, the application result is disappointing, which is caused by the failure of bubble generation in foam flooding. Therefore, it is necessary to seek ways for improving the performance of foam injection. An increased disturbance to the stratum, like the vibration caused by a seismic o...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Circulation journal : official journal of the Japanese Circulation Society
دوره 78 12 شماره
صفحات -
تاریخ انتشار 2014