Agreement of Central Corneal Thickness Measurements Using two Different Ultra Sound Techniques

نویسندگان

  • Dr Karthika Mohankumar
  • Rashmi Jain
  • Vidya Hegde
چکیده

Purpose: To compare the Central Corneal Thickness (CCT) measurements obtained by Ultrasound Pachymeter (USP) & Ultrasound Biomicroscope (UBM). Methods: The study CCT measurements of 60 eyes was measured using first with UBM followed by USP. Results were compared statistically using Bland –Altman plots & Deming’s regression analysis Results: Scatter plot analysis showed that for the same value of CCT UBM , corresponding CCTP measurements vary systematically. Bland – Altman plot analysis showed that there is systematic difference between the measurements with both instruments. Deming’s regression analysis confirmed a reasonable agreement between both two measurements. Conclusion: There is a positive correlation in our study & the readings obtained by UBM correlate with that measured by Pachymeter. But since the interobserver variability is high with UBM, pachymeter is considered to be a better modality of measurement of CCT among the two of them.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Measurements of central corneal thickness using two immersion ultrasound techniques and optical technique.

OBJECTIVE To compare the accuracy of central corneal thickness measurements using ultrasound biomicroscopy, Orbscan II tomography and an Artemis-2 very high frequency ultrasound scanner. METHODS The prospective study was conducted at Eye World Centre, Riyadh, Saudi Arabia, from September to November 2012. One eye from each of 60 normal subjects was analysed. The central corneal thickness was ...

متن کامل

Intraocular Pressure Measurements by Three Different Tonometers in Children with Aphakic Glaucoma and a Thick Cornea

Background: To evaluate the agreement in intraocular pressure (IOP) measurements by Ocular Response Analyzer (ORA) and Tono-Pen XL (TXL) with the Goldmann Applanation Tonometer (GAT) and to examine corneal biomechanical properties in aphakic glaucoma patients with a central corneal thickness (CCT) >600 µ. Methods: Thirty-six eyes of aphakic glaucoma patients (group 1) and 40 eyes of normal chil...

متن کامل

Comparison of Central Corneal Thickness Measured through Pentacam, OCTCasia2, and Orbscan IIz in Normal Individuals

Purpose: To compare the central corneal thickness measured through Pentacam, OCTCasia2, and Orbscan IIz in normal individuals. Methods: In this cross sectional comparative study, the researcher investigated the eyes of 70 normal candidates (140 eyes) with the mean age of 31.42 ± 7.70, referring to Dr. Sedaghat eye clinic for refractive surgery. For each person, Central Corneal Thickness (CCT) w...

متن کامل

Effect of corneal parameters on measurements using the pulsatile ocular blood flow tonograph and Goldmann applanation tonometer.

AIMS To investigate the effect of central corneal thickness and corneal curvature on intraocular pressure measurements using the pulsatile ocular blood flow tonograph and the Goldmann applanation tonometer, and to assess the agreement between the pulsatile ocular blood flow tonograph and the Goldmann applanation tonometer in intraocular pressure measurement. METHODS 479 subjects underwent int...

متن کامل

Comparison of central corneal thickness estimated by an ultrasonic pachymeter and non-contact specular microscopy.

Purpose: To compare central corneal thickness (CCT) measurements of healthy individuals obtained with ultrasonic pachymetry (UP) and non-contact specular microscopy (NCSM). Method: In total, 148 eyes of 74 subjects with no ocular or systemic diseases were included in the study. Central corneal thickness measurements of all patients performed with UP and NCCM were compared. Results: A total ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2016