Patient Dose-Area Product (DAP) Monitoring in Digital Radiography
نویسنده
چکیده
The objective of this study was to determine the Dose Area Product (DAP) for sample of 70 patients (40 Females and 30 Males) undergoing different types of digital radiography (DR) examinations at King Abdul-Aziz University Hospital (KAUH), Saudi Arabia. This study was done through implementing a quality control (QC) program initiated by The Center for Training and radiation protection, KAU. It was found that, the majority of the examinations carried out at the university hospital were Chest (46%), lumbar spine (13%), knee (10%), wrist lateral (6%) projection, and the rest for the other projections. For the most frequent value in this survey (Chest PA) the measured dose area product ranged from 0.12 to 0.42 Gy.cm2 with an average value of 0.24 Gy.cm2, for Chest lateral, it ranged from 0.18 to 1.48 Gy.cm2 with an average value of 0.65 Gy.cm2. For Cervical spine the observed DAP values ranged from 0.1 to 0.38 Gy.cm2 with an average value of 0.23 Gy.cm2. For Lumbar spine the value of the DAP ranged from 2.14 Gy.cm2 to 2.93 Gy.cm2 with an average value of 2.44 Gy.cm2. The highest average DAP value (2.44 Gy.cm2) was obtained from lumbar spine projections. The results indicate that, the results of this survey were within the corresponding international values available from the literature. Some additional measures needed to reduce the dose without loss of image quality especially for lumbar spine protocol. Monte Carlo technique was used to calculate the conversion factors that estimate the effective dose used for measuring the DAP or entrance skin dose.
منابع مشابه
A Comparison of Patients Absorption Doses with Bone Deformity Due to the EOS Imaging and Digital Radiology
Background: This study has aimed to measure the patient dose in entire spine radiography by EOS system in comparison with the digital radiography. Methods: EOS stereo-radiography was used for frontal and lateral view spine imaging in 41 patients in a prospective analytical study. A calibrated dose area product (DAP) meter was used for calibration of the DAP in EOS system. The accuracy and pr...
متن کاملImage quality assessment in digital radiography systems- A Pehamed FLUORAD A+D phantom study
Introduction: The use of digital radiography systems has become more common all around the world. Ideal situation in radiography is to maximize the image quality while minimizing the patient dose. The purpose of this study was to compare image quality and radiation dose in different digital radiography systems. Materials and Methods: Images of Pehamed FLUORAD ...
متن کاملA Phantom Study for the Optimization of Image Quality and Radiation Dose for Common Radiographic Examinations in Digital Radiography
Introduction: Phantom studies facilitate the implementation of radiation dose surveillance as a function of radiographic technical parameters for minimizing patient radiation dose. The evidence of such investigations can then be used to evaluate technical parameters used in the radiographic procedures to reduce radiation dose without compromising the image quality. Material and Methods: This ex...
متن کاملA Comparison of Patients Absorption Doses with Bone Deformity Due to the EOS Imaging and Digital Radiology.
BACKGROUND This study has aimed to measure the patient dose in entire spine radiography by EOS system in comparison with the digital radiography. METHODS EOS stereo-radiography was used for frontal and lateral view spine imaging in 41 patients in a prospective analytical study. A calibrated dose area product (DAP) meter was used for calibration of the DAP in EOS system. The accuracy and preci...
متن کاملThe Practice of Chest Radiography Using Different Digital Imaging Systems: Dose and Image Quality
Introduction: The study was undertaken to evaluate the practice of chest radiography using different digital imaging systems and its influence on dose and image quality. Materials and Methods: The study was carried out in two hospitals from March 2016 to June 2016. Sixty ambulatory patients aged 21 to 60 years who were able to cooperate without difficulty and weighed between 60 to 80 kg were ...
متن کامل